17

18

5) According to the wire connection interface (Fig. 10), clamp L3 live wire, connect L3 live wire 
    with test lead from V terminal, and connect L2 live wire with test lead from COM terminal, 
    then press “HOLD” button to enter P2 measurement interface, as shown below:

The finger symbol on “HOLD” segment flashes to indicate pressing “HOLD” button to enter 
phase-combined active power interface.

6) In P2 measurement interface, press “HOLD” button to read phase-combined active power, 
    press SELECT button to cyclically read phase-combined active power, P1 active power and 
    P2 active power, as shown below:

Second wire connection interface
Main display: Active power P2
Auxiliary display 1: P2 prompt
Auxiliary display 2: Voltage
Auxiliary display 3: Current

Second wire connection interface
Main display: Active power P1
Auxiliary display 1: P1 prompt
Auxiliary display 2: Voltage
Auxiliary display 3: Current

Phase-combined interface
Main display: Phase-combined active power P1+P2
Auxiliary display 1: Phase-combined prompt

Short press

4.4 3-phase 4-wire (3P4W) unbalance mode (Fig. 11)

L1

/

R

/

A

L2

/

S

/

B

N

L3

/

T

/

C

Red

Black

Power supply

AC 3-phase 4-wire power measurement (3P4W unbalance)

Load

Fig. 11

UT219P User Manual

UT219P User Manual

1) Connect red test lead (connected with red alligator clip already) with “V” terminal; and 
    black (connected with black alligator clip already) with “COM”.

2) Set the rotary switch to “3PW”, press “

Δ

/

” to select 3P4W unbalance mode, then the 

    clamp enters P1 wire connection interface, as shown below:

3P4W

“L1” on top right corner indicates the jaws clamp L1 live wire, “n” on bottom left corner 
indicates test lead from COM terminal connects with neutral wire, “L1” on bottom right corner 
indicates test lead from V terminal connects with L1 live wire. The finger symbol on “HOLD” 
segment flashes to indicate pressing “HOLD” button to enter P1 measurement interface.

3) According to the wire connection interface (Fig. 11), clamp L1 live wire, connect L1 live wire 
    with test lead from V terminal, and connect neutral wire with test lead from COM terminal, 
    then press “HOLD” button to enter P1 measurement interface, as shown below:

The finger symbol on “HOLD” segment flashes to indicate pressing “HOLD” button to enter 
P2 wire connection interface.

4) In P1 measurement interface, after the displayed value stabilizes, press “HOLD” button to 
    enter P2 wire connection interface, as shown below:

“L2” on top right corner indicates the jaws clamp L2 live wire, “n” on bottom left corner 
indicates test lead from COM terminal connects with neutral wire, “L2” on bottom right corner 
indicates test lead from V terminal connects with L2 live wire. The finger symbol on “HOLD” 
segment flashes to indicate pressing “HOLD” button to enter P2 measurement interface.

Summary of Contents for UT219P

Page 1: ...P N 110401111507X UT219P AC Digital Power Clamp Meter User Manual...

Page 2: ...end guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misuse...

Page 3: ...1000V AC 4 1000V AC RMS voltage measurement 15Hz 1000Hz frequency response range 5 1000A AC RMS current measurement 15Hz 1000Hz frequency response range 6 Ability to measure and display RMS and peak...

Page 4: ...uracy Remove the battery if the product is not used for a long time 10 Do not alter the internal wiring to avoid product damage and safety hazard 11 Do not keep or use the product in environments with...

Page 5: ...ctor 4 3P3W 3 phase 3 wire measurement mode 5 3P4W 3 phase 4 wire measurement mode 6 UNBALANCED Unbalanced measurement mode 7 MAX Maximum value measurement 8 MIN Minimum value measurement 9 AVG Averag...

Page 6: ...ximum value minimum value average value on the main display and to record the time of capturing maximum and minimum values Long press this button to exit maximum minimum value measurement The measurem...

Page 7: ...ip already with V terminal and black connected with black alligator clip already with COM 2 Set the rotary switch to A V press SELECT to select AC voltage measurement and connect the alligator clips w...

Page 8: ...lamp jaws are correct L1 R A L2 S B L3 T C Red Black Power supply AC 3 phase 3 wire power measurement 3P3W balance Load 4 3 phase power measurement 4 1 3 phase 3 wire 3P3W balanced mode Fig 8 Fig 8 UT...

Page 9: ...voltage Auxiliary display 3 AC current Short press Short press Short press Short press Short press 4 3 3 phase 3 wire 3P3W unbalance mode Fig 10 L1 R A L2 S B L3 T C Red Black Power supply AC 3 phase...

Page 10: ...alligator clip already with V terminal and black connected with black alligator clip already with COM 2 Set the rotary switch to 3PW press to select 3P4W unbalance mode then the clamp enters P1 wire c...

Page 11: ...ower Auxiliary display 1 P1 prompt Auxiliary display 2 Voltage Auxiliary display 3 Current Main display Second active power Auxiliary display 1 P2 prompt Auxiliary display 2 Voltage Auxiliary display...

Page 12: ...at the center of the clamp jaws Error of 1 0 of reading will be produced if the measured conductor is not placed at the center of the clamp jaws If zero crossing of harmonic does not meet the zero cr...

Page 13: ...tronic meter Press Hold to start accumulation when the LED flashes for one time press HOLD again to stop accumulation when the LED flashes again as shown below Electronic Meter LED flashes for one tim...

Page 14: ...ace of wire connection Fig 13 within 10 seconds connect test lead from V terminal with L3 live wire and connect test lead from COM terminal with L2 live wire then three kinds of testing results will b...

Page 15: ...al error of temperature coefficient is 0 1 specified accuracy C To ensure measurement accuracy please place the measured conductor at the center of the clamp jaws Error of 1 0 of reading will be produ...

Page 16: ...angle error 3 5 Phase angle 15Hz 40Hz 70Hz 1000Hz 40Hz 70Hz 1000A AC 1000Vrms 0 1 3 5 5 Range Resolution Accuracy Overload protection Remark Min measured voltage 30V Max measured current 10A 180 adva...

Page 17: ...ment result viewing button control and other operations can be achieved through UNI T Smart Measurement APP 3 2 UNI T Smart Measurement APP has multiple functions including Bluetooth communication dat...

Reviews: