background image

16

15

    “OL” is displayed if the measured diode is open or the polarity is reversed.
    Before measuring on-line diode, please switch off all powers of measured 

circuit, and discharge all capacitors completely.

    The open-circuit voltage for diode measurement is about 3.0V.
    Do not input voltage over 30V (DC/AC) to avoid personal injury.
    Disconnect the test lead with measured circuit after all measurement 

operations are completed.

7. Capacitance measurement (Figure 8)

1) Set the rotary switch to “         ”, then short press SEL to select capacitance 
    measurement.
2) Connect red test lead with V terminal, and black with COM. Then connect 
    test leads with both ends of measured capacitor.
3) Read the measured capacitance from the LCD. 

  Warning:

   It is recommended to measure capacitor below 100nF in REL mode.
   “OL” is displayed on the LCD if the measured capacitor is shorted or the 
   measured capacitance is over the maximum range.
   Please discharge capacitor completely before measurement (especially for 
   capacitor with high voltage) to avoid product damage or personal injury.
   Disconnect the test lead with measured circuit after all measurement 
   operations are completed.

8. Current probe measurement (ACA/DCA) (Figure 9)

1) Connect current probe with V terminal and COM terminal.
2) Set the rotary switch to “          ”, then press SEL to select ACA/DCA 
    measurement.
3) Read from the LCD the measured current of current probe. Under ACA 
    mode of current probe, short press SEL to switch to frequency measurement 
    in ACA mode of current probe, then read from the LCD the frequency of 
    current.

 Warning:

    For frequency measurement under ACA mode of current probe, the input 

current amplitude shall be greater than 10% of full range.

    The frequency response is 45Hz~400Hz. For range at 600.0A: 1mV = 1A 

(AC/DC). The current probe measurement is an input mode of analog voltage, 
the specified accuracy and frequency response refer to the intrinsic accuracy 
and frequency response of UT117C (eliminate the error of current probe).

    Disconnect the test lead with measured circuit after all measurement 

operations are completed.

9. AC/DC current measurement (Figure 9)

1) Set the rotary switch to “      ”/“        ”.
2) Connect red test lead with A terminal, and black with COM. Then connect 
    test lead with measured loop in series.
3) Read the measured current from the LCD. The displayed AC current is true 
    RMS value.
4) When measuring AC current, short press SEL to switch to frequency 
    measurement under AC current mode, then read from the LCD the frequency 
    of measured current. For measuring frequency of current, the AC current 
    amplitude shall be 600mA to 10A; the input amplitude shall be greater than 
    3A for frequency measurement at range of 10A. If the measured frequency 
    is unstable, please select low current range manually for a stable readout. 
    See “Technical Specifications” for details.
5) For frequency detection, analog bar graph and range indicator indicate the 
    current AC current.

Warning:

    

When turning the rotary switch to or from “         ”/“        ”, the LCD shows 

    

“LEAd” for about a second as a warning of testing lead.

    

Please check if the fuse is good before measuring current. The instructions 

    

to check the fuse is as below: 

    

Set the rotary switch to Ω position, then short-circuit V terminal and A 

    

terminal. If the displayed resistance is about 0.0Ω, the fuse is good, 

    

otherwise damaged.

UT117C User Manual

UT117C User Manual

Figure 8

Figure 9

Summary of Contents for UT117C

Page 1: ...UT117C P N 110401111679X High Precision True RMS Digital Multimeter User Manual...

Page 2: ...end guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misuse...

Page 3: ...d auto power off and Bluetooth II Features True RMS to ensure accurate measurement of non linearity load The ability to measure 20A 10 seconds of transient measurement LoZ function Low impedance to pr...

Page 4: ...asured devices and discharge all capacitors completely Before measuring current please check if the fuse of the multimeter is good and switch off the measured current before connecting the multimeter...

Page 5: ...IEC 61010 1 CAT 600V VII External Structure 1 NCV sensing end 2 LCD display 3 Functional buttons Used to select measurement functions 4 Rotary switch 5 A terminal 6 COM terminal 7 V terminal 8 Holder...

Page 6: ...our times at a frequency of 2Hz and the auto off function is disabled To restore auto range and auto off function please exit MAX MIN mode If the auto off function is disabled manually the multimeter...

Page 7: ...ction Composite sinusoidal signal generated by inverter or variable frequency motor can be measured through the LPF function as shown in Figure 6 below Long press SELECT again to exit LPF function 5 W...

Page 8: ...ent Do not input voltage over 30V DC AC to avoid personal injury Disconnect the test lead with measured circuit after all measurement operations are completed 5 Resistance measurement Figure 6 1 Set t...

Page 9: ...asurement under ACA mode of current probe the input current amplitude shall be greater than 10 of full range The frequency response is 45Hz 400Hz For range at 600 0A 1mV 1A AC DC The current probe mea...

Page 10: ...V AC please observe if the conductor of measured electric field is insulated so as to avoid personal injury Even the displayed result is that voltage is not present it does not mean voltage is present...

Page 11: ...ution Accuracy a of reading b digit Overload protection Input impedance About 10M Range to ensure accuracy 1 100 of range Least significant digit under short circuit 2 digits 600 00mV 0 01mV 1 30 600V...

Page 12: ...racy 1 100 of range The least significant digit at current range under open circuit is allowed to be 2 Note 10 20 A Perform measurement continuously for 10 seconds at most then stop for 15 minutes at...

Page 13: ...allowed to be about 5 under short circuit 12 Current probe measurement ACA DCA 600 0A 0 1A 1 8 3 600V AC DC 600 0A 0 1A 0 8 2 ACA DCA Range Resolution Accuracy a of reading b digit Overload protectio...

Page 14: ...a synchronizing and more Please refer to the use manual of iDMM2 0 for use of these functions 4 Uninstallation of the software Uninstall the software through the uninstallation function of mobile phon...

Page 15: ...o not keep in environments with high humidity high temperature and strong electromagnetic fields 2 Replace battery or fuse Figure 11 Warning 1 When the symbol shows on the LCD please replace the batte...

Reviews: