background image

14

13

  Warning:

    The input impedance of the multimeter is about 10MΩ. Measurement error 

will be produced when measuring circuit with high impedance. In most 
cases, the impedance of circuit is below 10KΩ, thus error of 0.1% or less 
can be negligible.

    Do not measure input voltage over the range, otherwise correct reading 

cannot be obtained and product damage and personal injury may occur.

    Do not input voltage over 600.0mV. It is possible to measure higher voltage, 

but the protection provided by the multimeter can be compromised.

    Pay special attention to avoid electric shock when working with high voltage.
    To check if the multimeter can function well, please measure a known 

voltage before use.

    If the measured voltage is >600V (AC/DC), the high voltage symbol “     ” 

appears.

    Disconnect the test lead with measured circuit after all measurement 

operations are completed.

4. Continuity measurement (Figure 6)

1) Set the rotary switch to “        ”, then short press SEL to switch to continuity 
    (      ) measurement.
2) Connect red test lead with V terminal, and black with COM. Then connect 
    test lead with both ends of load of measured circuit in parallel.
3) Read from the LCD the resistance of load of measured circuit.

Warning:

    

Audible and visual alarm (not in silent mode): If the resistance between both 
measured ends is ≤20Ω, the buzzer makes a long beep and the green light 
is lit up; if >20Ω, the red light is lit up and the OL light is off.

    

Before measuring on-line continuity, please switch off all powers of measured 
circuit, and discharge all capacitors completely.

    

The open-circuit voltage is about 2V for continuity measurement.

    

Do not input voltage over 30V (DC/AC) to avoid personal injury.

    

Disconnect the test lead with measured circuit after all measurement 

operations are completed.

5. Resistance measurement (Figure 6)

1) Set the rotary switch to “        ”, then short press SEL to switch to resistance 

(Ω) measurement.

2) Connect red test lead with V terminal, and black with COM. Then connect 

test leads with both ends of measured resistor in parallel.

3) Read the measured resistance from the LCD.

  Warning:

    

“OL” will be displayed on the LCD if the measured resistor is open or the 

    

measured resistance is over the maximum range.

    

Before measuring on-line resistance, please switch off all powers of 

    

measured circuit, and discharge all capacitors completely.

    

If the resistance of shorted test lead is ≥0.5Ω, please check if test lead is 

    

loose or other problems occur.

    

It is normal to take several seconds to stabilize the reading if the measured 

    

resistance is over 1MΩ.

    

Do not input voltage over 30V (DC/AC) to avoid personal injury.

    

Disconnect the test lead with measured circuit after all measurement 

    

operations are completed.

6. Diode measurement (Figure 7)

1) Set the rotary switch to “            ”, then short press SEL to select diode 
    measurement.
2) Connect red test lead with V terminal, and black with COM.  The polarity 

of red test lead is “+”, and black is “-”. Connect red test lead with positive 
polarity of diode, and black with negative.
3) Read from the LCD the approximate forward voltage of PN junction of 
measured diode. The normal voltage of silicon PN junction is about 
0.5~0.8V.

Warning:

    Audible and visual alarm (not in silent mode): For < 0.12V, the red light is 

lit up and the buzzer makes a long beep; for 

0.12V and < 2V, the green 

light is lit up and the buzzer sounds once; for >2V, the light is off.

UT117C User Manual

UT117C User Manual

Figure 6

Figure 7

Summary of Contents for UT117C

Page 1: ...UT117C P N 110401111679X High Precision True RMS Digital Multimeter User Manual...

Page 2: ...end guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misuse...

Page 3: ...d auto power off and Bluetooth II Features True RMS to ensure accurate measurement of non linearity load The ability to measure 20A 10 seconds of transient measurement LoZ function Low impedance to pr...

Page 4: ...asured devices and discharge all capacitors completely Before measuring current please check if the fuse of the multimeter is good and switch off the measured current before connecting the multimeter...

Page 5: ...IEC 61010 1 CAT 600V VII External Structure 1 NCV sensing end 2 LCD display 3 Functional buttons Used to select measurement functions 4 Rotary switch 5 A terminal 6 COM terminal 7 V terminal 8 Holder...

Page 6: ...our times at a frequency of 2Hz and the auto off function is disabled To restore auto range and auto off function please exit MAX MIN mode If the auto off function is disabled manually the multimeter...

Page 7: ...ction Composite sinusoidal signal generated by inverter or variable frequency motor can be measured through the LPF function as shown in Figure 6 below Long press SELECT again to exit LPF function 5 W...

Page 8: ...ent Do not input voltage over 30V DC AC to avoid personal injury Disconnect the test lead with measured circuit after all measurement operations are completed 5 Resistance measurement Figure 6 1 Set t...

Page 9: ...asurement under ACA mode of current probe the input current amplitude shall be greater than 10 of full range The frequency response is 45Hz 400Hz For range at 600 0A 1mV 1A AC DC The current probe mea...

Page 10: ...V AC please observe if the conductor of measured electric field is insulated so as to avoid personal injury Even the displayed result is that voltage is not present it does not mean voltage is present...

Page 11: ...ution Accuracy a of reading b digit Overload protection Input impedance About 10M Range to ensure accuracy 1 100 of range Least significant digit under short circuit 2 digits 600 00mV 0 01mV 1 30 600V...

Page 12: ...racy 1 100 of range The least significant digit at current range under open circuit is allowed to be 2 Note 10 20 A Perform measurement continuously for 10 seconds at most then stop for 15 minutes at...

Page 13: ...allowed to be about 5 under short circuit 12 Current probe measurement ACA DCA 600 0A 0 1A 1 8 3 600V AC DC 600 0A 0 1A 0 8 2 ACA DCA Range Resolution Accuracy a of reading b digit Overload protectio...

Page 14: ...a synchronizing and more Please refer to the use manual of iDMM2 0 for use of these functions 4 Uninstallation of the software Uninstall the software through the uninstallation function of mobile phon...

Page 15: ...o not keep in environments with high humidity high temperature and strong electromagnetic fields 2 Replace battery or fuse Figure 11 Warning 1 When the symbol shows on the LCD please replace the batte...

Reviews: