7. Operating Basics 

 

UT-P35&UT-P36 high voltage isolating probe

 is very convenient 

to operate. But we suggest every new user should read the 

manual perfectly prior to begin operating. 

 

7-1 Operation steps 

a.

 

Check and affirm the AC adapter input voltage is correct prior to 
use. 

b.

 

Guesstimate the voltage to measure and set the correct attenuation 
switch. If the input voltage is over range, perhaps it will damage the 
probe. 

c.

 

Connect the BNC output terminal to oscilloscope or other measuring 
instrument. Connect the input cables to the adapt clips well.   

d.

 

Connect the AC adapter output to probe power input connector. The 
green POWER indicator lights. 

e.

 

Select adapt attenuation range according to measured voltage. When 
the measured voltage is over range, the red OVER RANGE 
indicator lights, the attenuation should increase or disconnect the 
power and the input measured voltage. 

f.

 

Set the measuring oscilloscope or other instrument sensitivity 
according to displayed altitude of signal. 

g.

 

Connect the adapt clips to input cables. 

h.

 

Connect the clips to measured targets to start measuring. The probe 
body should be far from high voltage pulse circuits to reduce the 
noise or errors. 

i.

 

When the test is finished, disconnect the power at first, then 
disconnect input from measured target, disconnect the output BNC 
terminal from the oscilloscope or other measuring instrument at last.

   

Summary of Contents for UT-P35

Page 1: ...UT P35 UT P36 High Voltage Isolating Probes Instruction Manual...

Page 2: ...1 Please read the instruction manual carefully to avoid injury and prevent damage to this product Ver 1 0 2021 05 19...

Page 3: ...lacement in exchange for the defective product Parts modules and replacement products used by Instrance for warranty work may be new or reconditioned to like new performance All replaced parts modules...

Page 4: ...able of contents 1 Safety Summary 4 2 Brief Introduction 5 3 Main Specification 6 4 Unpacking and Preparation for Use 8 5 General View of Product 9 6 General View of Accessories 10 7 Operating Basics...

Page 5: ...sconnect probes while they are connected to a voltage source Ground the product This product is indirectly grounded through the grounding conductor of the mainframe power cord To avoid electric shock...

Page 6: ...C peak AC accordingly UT P35 UT P36 high voltage isolating probe has over range alarm function It indicates with a red LED UT P35 UT P36 high voltage isolating probe is small and robust It has very hi...

Page 7: ...uts 8M Input capacitance Between each input to ground 7pF Between inputs 3 5pF Maximum measurable differential voltage 1 50 130V DC peak AC 1 500 1300V DC peak AC Maximum input voltage to earth 1000Vr...

Page 8: ...P35 60cm UT P36 30cm Output cable length 90cm Alligator clips 85 40 13 mm Pincer clips 160 45 13 mm Probe body 120 55 24 mm Weight probe only 160g Environment Specification Operating temperature 0 40...

Page 9: ...Instruction manual 1pcs AC adapter 1pcs Red alligator clip 1pcs Black alligator clip 1pcs Red pincer clip 1pcs Black pincer clip 1pcs 4 2 Preparation for use Please check the line voltage prior to con...

Page 10: ...9 5 General View of Product...

Page 11: ...View of Accessories UT P35 UT P36 high voltage isolating probe includes following accessories An AC adapter a couple of alligator clips and a couple of pincer clips 1 AC adapter 2 Alligator clips 3 P...

Page 12: ...probe power input connector The green POWER indicator lights e Select adapt attenuation range according to measured voltage When the measured voltage is over range the red OVER RANGE indicator lights...

Page 13: ...ld be lower than 10MHz Otherwise the errors will be larger c The probe and measuring instrument should warm up for 20 minutes if want to get high accuracy measuring result 7 3 Troubleshooting The Powe...

Page 14: ...13...

Reviews: