UDP6720 Series Digital Control Power Supply
User’s Manual
16
About
1) Push
to enter the setup interface;
2) Push
or
to select About field;
3) The device information product’s model and the current version number will shown on the right side of the
screen.
Remote Sense Compensation Setup
Before use Remote Sense compensation function, take off short connect terminal on the Remote Sense compensation
terminal, connect load voltage positive terminal to compensation terminal S+ and load voltage negative terminal connect
to compensation terminal S-. And then enable power output to turn on Remote Sense compensation function. That is,
the output voltage of the DC power supply is the voltage of the load terminal.
Notice: Remote Sense compensation voltage should less than 0.6V.