Programming Manual UDP3000S Series
Instruments.uni-trend.com
37
/
49
AUTO
:
Automatic trigger, the instrument can always generate the trigger condition at this time; it cannot have
parameter with<value>
;
OUTOFF
:
It will generate the trigger condition when the controlled source turn off the ouput; It cannot have
parameter with<value>
;
OUTON
:
It will generate the trigger condition when the controlled source turn on the ouput
;
It cannot have
parameter with<value>
;
>V
:
It will generate the trigger condition when the voltage of the controller source is greater than the
judgement value; The parameter <value> cannot be omitted
;
<V
:
It will generate the trigger condition when the voltage of the controller source is less than the judgement
value; The parameter <value> cannot be omitted
;
=V
:
It will generate the trigger condition when the voltage of the controller source is equal to the judgement
value
;
The parameter <value> cannot be omitted
;
>C
:
It will generate the trigger condition when the current of the controller source is greater than the judgement
value; The parameter <value> cannot be omitted
;
<C
:
It will generate the trigger condition when the current of the controller source is greater than the judgement
value; The parameter <value> cannot be omitted
;
=C
:
It will generate the trigger condition when the current of the controller source is equal to the judgement
value; The parameter <value> cannot be omitted
;
>P
:
It will generate the trigger condition when the power of the controller source is greater than the judgement
value; The parameter <value> cannot be omitted
;
<P
:
It will generate the trigger condition when the power of the controller source is less than the judgement
value; The parameter <value> cannot be omitted
;
=P
:
It will generate the trigger condition when the power of the controller source is equal to the judgement
value
;
The parameter <value> cannot be omitted.
For Example
:
:TRIGger:OUT:CONDition D0,>V,30.00 Set the trigger condition of IO1 to the voltage is greater
than 30.00 volts.
:TRIGger:OUT:CONDition
Format
:
:TRIGger:OUT:CONDition {D0|D1|D2|D3}
Function
:
Query the trigger condition of output mode trigger.
Description
:
D0|D1|D2|D3
:
Select output IO
,
D0-IO1
,
D1-IO2
,
D2-IO3
,
D3-IO4.
Return
:
“{AUTO|OUTOFF|OUTON|>V|<V|=V|>C|<C|=C|>P|<P|=P} [,<value>]”
。
Return“AUTO|OUTOFF|OUTON”, it cannot have parameter with<value>
;
Return“>V|<V|=V|>C|<C|=C|>P|<P|=P”, it have parameter<value>.
For Example
:
:TRIGger:OUT:CONDition? D0
Query the output trigger condition of IO1.
Return
:
>V,30.00
Return result
:
trigger condition is voltage is greater than
30.00 volts.
:TRIGger:OUT:POLArity
Format
:
:TRIGger:OUT:POLArity {D0|D1|D2|D3},{POSItive|NEGAtive}