Programming Manual UDP3000S Series
Instruments.uni-trend.com
19
/
49
Function
:
Query the actual voltage, current and power value at the output terminal of the specified channel.
Description
:
If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default.
For Example
:
:MEASure:ALL? CH1
Return
:
05.10,0.089,00.45
:MEASure[:VOLTage][:DC]?
Format
:
:MEASure[:VOLTage][:DC]? [CH1|CH2|CH3|SER|PARA]
Function
:
Query the actual voltage value at the output terminal of the specified channel.
Description
:
If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default.
For Example
:
:MEASure:VOLTage? CH1
Return
:
05.10
:MEASure:CURRent[:DC]?
Format
:
:MEASure:CURRent[:DC]? [CH1|CH2|CH3|SER|PARA]
Function
:
Query the actual current value at the output terminal of the specified channel.
Description
:
If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default.
For Example
:
:MEASure:CURRent? CH1
Return
:
0.089
:MEASure:POWEr[:DC]?
Format
:
:MEASure:POWEr[:DC]? [CH1|CH2|CH3|SER|PARA]
Function
:
Query the actual power value at the output terminal of the specified channel.
Description
:
If the channel parameter is ignored, then the current channel which is the channel need to be
queried by default.
For Example
:
:MEASure:POWEr? CH1
Return
:
00.45
7. LISTout
:LISTout[:STATe]
Format
:
:LISTout[:STATe] {0|1|OFF|ON}
Function
:
Start or stop the list output mode of the current channel.
Description
:
It can set and query the current channel by the command in“INSTrument”chapter.
For Example
:
:LISTout:STATe ON