UNDATEC FDL400 Manual Download Page 4

 

UNDATEC Flow Measurement Solutions

Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - T49 (0) 7237-7804
Email: [email protected]

 

 

2. Safety instructions

 

Safe operation is only ensured if the notes and warnings will be noted in this manual. 

 

Important! Seal, seals and labels

 

Opening or removing seals or labels, eg etc. with serial numbers, will result in immediate loss 
of warranty claims result.  
 

Determination and proper use 

The manufacturer is not liable for damages resulting from improper use. 

Compare and information before operating the conformity of the supply voltage match the 
specifications on the nameplate. 

If it becomes that a safe operation is no longer possible (eg visible damage), put the appliance 
out of service immediately and secure it against unintentional starting. 

Improper or unauthorized use, dangers may arise from the device, which is why we refer to 
consistent observance of safety notes. 

 

Mounting, commissioning and operating personnel

 

Mounting, electrical installation, commissioning and maintenance of the device must be carried out only 
by trained specialist personnel authorized by the plant operator. 

Technical personnel must have read and understood this manual, and follow their statements. 

The device may only be operated by persons who are authorized and trained by the facility operator. The 
instructions in this manual must be followed. 

Make sure that the device is in accordance with the electrical terminals connected correctly.

 

 

Repairs

 

Repairs should only be performed by trained service personnel. 

Please contact in this case to the company Dömges.Detectronic UG.

 

 

Technical progress

 

The manufacturer reserves the right to modify technical data without special announcements on the basis 
of development of technical progress. About the activities and possible extensions of this manual can be 
supplied by the company Dömges Detectronic UG. 

 

 

 

Summary of Contents for FDL400

Page 1: ...UNDATEC Flow Measurement Solutions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de FDL400 Flow Data Logger...

Page 2: ...2 Safety instructions intended use 3 Description 4 FDL400 4 1 Specifications 4 2 Dimensions 4 3 Pin Assignment 4 4 Connection Diagram 5 Programming 6 USB port and SD card 6 1 Reading data from the SD...

Page 3: ...d product with the provisions of the Directive evidenced by full compliance with the following harmonized European standards 2014 30 EU EMC Directive DIN EN 61000 6 2 2006 Corrigendum 1 2011 06 DIN EN...

Page 4: ...from the device which is why we refer to consistent observance of safety notes Mounting commissioning and operating personnel Mounting electrical installation commissioning and maintenance of the devi...

Page 5: ...is collected and further processed in the processor The integrated pressure sensor is used to detect the level in m above this value each perfused segment surface is again in the processor through a p...

Page 6: ...own fuse must not be repaired by the customer since this is special parts for ATEX approved devices Protection class IP30 Dimension 197 x 153 x 37 5mm B x H x T Temperatur range 0 50 C Display LCD 4 l...

Page 7: ...UNDATEC Flow Measurement Solutions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de 4 2 Dimension...

Page 8: ...ions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de 4 3 Pin Assignment Attention In ATEX relevant applications never do operate the sensor wit...

Page 9: ...ail undatec web de 4 4 Connection Diagram The current outputs are not isolated from the power supply The switching outputs are electrically isolated by optocouplers from the device power supply and th...

Page 10: ...pecial function Functions Button B generated entry into the function menu Here the desired function will be selected by entering the corresponding function number The functions are designed so that th...

Page 11: ...ounter this accumulated the amount as long until it is reset Function 10 Calibration factor for velocity V software based amplification of the V value 0 001 9 999 Function 11 Calibration factor for le...

Page 12: ...e factory generated calibration value of the analog current and or voltage outputs can be displayed Function 80 Reset of Totalizer select Totalizer or Totalizer per day Function 90 Automatic zeroing V...

Page 13: ...D card Secure Digital Memory Card This is on top of the unit under an openable flap The presence and correct operation is indicated by lighting an LED to the side of SD card The saving of measuring da...

Page 14: ...low levels relatively linear flow rate curve This is necessary because of the velocity measuring sensor is able to measure up to a filling level of about 400mm safely An increasing velocity at higher...

Page 15: ...Email undatec web de 6 Top panel openable flap with SD card and USB connection Figure shows open flap on top of the device with SD card and USB connection The USB interface is neccessary for update th...

Page 16: ...is a space Example With the most spreadsheet programs it is possible to design curves graphs and statistical analysis FDL400 Logfile Date Time V m s H m Q l s M m3 2015 03 15 14 42 13 0 02 m s 12 mm...

Page 17: ...ovided Using the supplied bracket the unit can be mounted in the control cabinet using a DIN rail For this purpose the 3 pcs Screws with the rosettes on the back of the unit must be loosened With the...

Page 18: ...done as shown in the photo To engage the rail the device is so compressed the spring down This can now be folded down in the DIN rail in the unit To disassemble the device it is also compressed down...

Page 19: ...UNDATEC Flow Measurement Solutions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de 8 Type Examination Certificate...

Page 20: ...UNDATEC Flow Measurement Solutions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de...

Reviews: