5.    FDL400 Programming 
 

After switching on the device is displayed on the display device identification , hardware version, software 

version, and copyright. 

 

Example: 

 

             FDL400 

      H – Version 2.0 

      S – Version 2.3 

   copyright UNDATEC

 

 

After a few seconds the display will jump to the measurement menu. 

 
The display shows velocity, level, flow and totalizer. 

 

Example: 

 

Velocity       0.00   m/s 

Level            0.000    m 

Flow            0000.0 l/s 
Total          000000 x 1 m3

 

 

Function buttons

 

 

Button  „B”  =  Break (Interrupting the measuring program for inputting parameters) 

 
Button  „E“  =  Enter 

 

Button  „F“  =  Correction 

 

Button   „A“  =  Abbort 

 

Button   „C“ =  no function

 

 

 

Button   „D“ =  special function  

 

Functions 

(Button „B“ generated entry into the function menu. Here, the desired function will be 

selected by entering the corresponding function number)

 

 

The functions are designed so that they are self-explanatory based on previous selection.

 

 

 

Function  1:  

Entering of channel shape and associated parameters. It can be a pre-programmed 

                     channel shape 

                     (Circle, rectangle, trapezoid, triangle or special) are selected. 

                      If special channel shapes are available, they are read in via memory card (SD card). 
 

Function  2:  

Enter the maximum values for V, H and Q. 

                     (Maximum values refer to the 0/4 - 20mA current outputs. Maximum value is the value at 

                      which for the selected output the current of 20mA has the full scale). 

 

Function  3:  

Entering the limit values for V, H and Q.  

                     (Limits refer to the switch outputs 1,2 and 3. Switch output 4 is for a different   
                     Application allocates.  

                     Value is the value at which the selected switching output switches).  

 

 

 

 

Summary of Contents for FDL400

Page 1: ...UNDATEC Flow Measurement Solutions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de FDL400 Flow Data Logger...

Page 2: ...2 Safety instructions intended use 3 Description 4 FDL400 4 1 Specifications 4 2 Dimensions 4 3 Pin Assignment 4 4 Connection Diagram 5 Programming 6 USB port and SD card 6 1 Reading data from the SD...

Page 3: ...d product with the provisions of the Directive evidenced by full compliance with the following harmonized European standards 2014 30 EU EMC Directive DIN EN 61000 6 2 2006 Corrigendum 1 2011 06 DIN EN...

Page 4: ...from the device which is why we refer to consistent observance of safety notes Mounting commissioning and operating personnel Mounting electrical installation commissioning and maintenance of the devi...

Page 5: ...is collected and further processed in the processor The integrated pressure sensor is used to detect the level in m above this value each perfused segment surface is again in the processor through a p...

Page 6: ...own fuse must not be repaired by the customer since this is special parts for ATEX approved devices Protection class IP30 Dimension 197 x 153 x 37 5mm B x H x T Temperatur range 0 50 C Display LCD 4 l...

Page 7: ...UNDATEC Flow Measurement Solutions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de 4 2 Dimension...

Page 8: ...ions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de 4 3 Pin Assignment Attention In ATEX relevant applications never do operate the sensor wit...

Page 9: ...ail undatec web de 4 4 Connection Diagram The current outputs are not isolated from the power supply The switching outputs are electrically isolated by optocouplers from the device power supply and th...

Page 10: ...pecial function Functions Button B generated entry into the function menu Here the desired function will be selected by entering the corresponding function number The functions are designed so that th...

Page 11: ...ounter this accumulated the amount as long until it is reset Function 10 Calibration factor for velocity V software based amplification of the V value 0 001 9 999 Function 11 Calibration factor for le...

Page 12: ...e factory generated calibration value of the analog current and or voltage outputs can be displayed Function 80 Reset of Totalizer select Totalizer or Totalizer per day Function 90 Automatic zeroing V...

Page 13: ...D card Secure Digital Memory Card This is on top of the unit under an openable flap The presence and correct operation is indicated by lighting an LED to the side of SD card The saving of measuring da...

Page 14: ...low levels relatively linear flow rate curve This is necessary because of the velocity measuring sensor is able to measure up to a filling level of about 400mm safely An increasing velocity at higher...

Page 15: ...Email undatec web de 6 Top panel openable flap with SD card and USB connection Figure shows open flap on top of the device with SD card and USB connection The USB interface is neccessary for update th...

Page 16: ...is a space Example With the most spreadsheet programs it is possible to design curves graphs and statistical analysis FDL400 Logfile Date Time V m s H m Q l s M m3 2015 03 15 14 42 13 0 02 m s 12 mm...

Page 17: ...ovided Using the supplied bracket the unit can be mounted in the control cabinet using a DIN rail For this purpose the 3 pcs Screws with the rosettes on the back of the unit must be loosened With the...

Page 18: ...done as shown in the photo To engage the rail the device is so compressed the spring down This can now be folded down in the DIN rail in the unit To disassemble the device it is also compressed down...

Page 19: ...UNDATEC Flow Measurement Solutions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de 8 Type Examination Certificate...

Page 20: ...UNDATEC Flow Measurement Solutions Frank Stark Wiesenstrasse 23 D 75248 Oelbronn Duerrn Germany Telefon 49 0 7237 7804 Email undatec web de...

Reviews: