29
9.2 Calibration of HYPROP BALANCE Series
Applicable models: HYPROP BALANCE
※
Span adjustment should be performed with the balance installed perfectly level and without any
load on the measurement pan.
1
Invoke the span-adjustment function.
Hold down the Function key, then
release your finger when the display
switches from [Func] to [S.A. CAL].
2
Start span adjustment.
Press the Function key while holding
down the Zero/Tare key, then
release both fingers simultaneously.
Display will change to a flashing [CAL.
0] and the zero-point will be calibrated
automatically.
3
Zero-point calibration
Once zero-point calibration is
completed, display switches to [CAL.
on]. Calibration now proceeds to that
of the weighing capacity point.
4
Calibration of weighing capacity point
Slowly and fully turn the calibration
knob in the direction of [Calibration].
The balance assumes the state where
the calibration weight is loaded on the
measurement pan.
The display starts flashing and the
balance automatically adjusts the
weighing capacity point.
Hold down.
Release your
finger.
Release both
fingers at the
same time.
Calibration