SK00-9663-E8-003-00
Command
Description
Command
Description
D
Read measured value and
status
T
Reading the model and
software version
ZER
Zero point adjustment
1R
Read Setpoint1 value
ATM
Atmospheric pressure
adjustment
2R
Read Setpoint2 value
CLR
Clear the zero point and
atmospheric pressure
adjustment
1W
Write Setpoint1 value
SR
Reading status
2W
Write Setpoint2 value
SW
Writing status
Command (Excerpt): Reading measured value and status
Command
:
AD0
AD1
D
CHKH
CHKL
CR
Return format from this unit to PC
: AD0 AD1 D X . X X E ± X X SH SL CHKH CHKL CR
「
X.XXE±XX
」
is the measured pressure value.
e.g.1) 3.00E+03
⇒
3.00×10
+3
e.g.2) 5.00E+00
⇒
5.00×10
+0
e.g.3) 4.00E-01
⇒
4.00×10
-1
When
「
E.EEE+EE
」
is returned.: Sensor error
When
「
F.FFE+FF
」
is returned.: The measurement range is exceeded
or filament is off.
Please refer to Section4.3 for SH and SL.