3. Setting up
3.1 Installation
Installation must be done at a place of low moisture, low dust particle, and
also level. The layout arrangement should be considered for setting up, overhaul,
checking, or cleaning the pump. When setting up the pump in any systems, pay
attention to temperature in atmosphere. Mount rubber isolator to absorb
vibration from the pump. On the environmental condition, refer to 「0.4.2 Store
of product」.
Do not tilt, lay its side, or operate up side down. It may cause damage.
The pump must be placed on a level place in an upright position as shown
in Fig-1,Fig-2.
3.2 Oil charging
Remove the Exhaust port, and then fill the oil (one time charge) to the pump
or recommended oil (SMR-100) up to specified quantity through the Exhaust port.
In first oil charging, pour the oil up to maximum scale of oil gauge. Then, attach
the exhaust port. (Refer to Fig-3) Oil quantity must be checked or adjusted to
keep the oil quantity is always in the range of oil gauge during operating the
pump.
Improper oil quantity may cause the performance deterioration of the pump, and
also may cause damage. When oil quantity is out of gauge scale ultimate pressure
become higher, and it may cause bubbling sound like "poko poko".