UltraSonic LLC. Model Ultra 6000FLT   

                 

11 

 

Push 

Power OFF 

button (all control lights should be OFF). 

 

 

DISCONNECT UNIT FROM POWER SUPPLY.

 

 

 

Wait a minimum of 30 minutes before draining contaminated solution from the tanks 
(this enables heating element to cool and prevent damage) 

 

 

Drain both tanks, 

NOTE:  

Solution is hot and may be contaminated; protective 

clothing and equipment should be used to prevent any exposure. 

 

 

Remove and clean metal pre-filter located in rear tank. 

 

 

Rinse the inside of the tanks with clean, clear water. 

 

 

Wipe down the inside of tanks with a clean, soft cloth.  Do not use any abrasive pads 
or cleaners.  These kinds of cleaners can scratch the tank surface and void warranty. 

 

 

Wipe work and outside surfaces with a dry, clean, soft cloth. 
 

 

When discarding contaminated solution and filters follow all local, state and 
federal environmental and regulatory requirements. 

 

 

Follow installation instructions listed earlier for refilling and start up procedures. 

 

 
MAXIMIZING ULTRASONIC PERFORMANCE 
 

 

Temperature - Temperature has a dramatic effect on the cleaning performance and 
time needed to clean a part.  The recommended temperature range is 140

F to 150

F.  

Temperatures above can decrease the ultrasonic action.  Consideration needs to be 
given to the cleaning agent used when determining the proper temperature.  

NOTE:  

Always check the MSDS for the cleaning agent and note its boiling point or flash 
point.  Operation near or above the boiling point/flash point is hazardous and can 
cause noxious fumes. 

 

 

Cleaning Agent - Cleaning agents can enhance or decrease the performance of the 
ultrasonic action and speed of cleaning.  Always use UltraSonic LLC recommended 
cleaning agents at their proper concentrations. 

 

 

Degassing - Liquids containing air and dissolved gases can interfere with the 
formation of cavitation bubbles necessary for ultrasonic cleaning.  Follow the 
procedure in the installation section to ensure liquid is ready for cleaning.   

 

 

Part Exposure - In order for a surface to be cleaned, it must be exposed to the liquid.  

Summary of Contents for 6000FLT

Page 1: ...LT 220v 240V 3 Voltage Supply to Machine L1 to L2 ___________ L1 to Ground ____________ L2 to Ground ___________ UltraSonic LLC 1200 Ferris Road Amelia Ohio 45102 Tel 513 502 9746 Fax 513 436 1527 inf...

Page 2: ...on Never attempt to perform maintenance when cleaning solution is hot Never open equipment or attempt service warranty related repairs without first contacting UltraSonic LLC for safe and proper proce...

Page 3: ...in the tank longer than necessary This can result in erosion of the part surface and or the bottom of the tank Do not allow moisture to build up inside the unit this can cause electrical shorts and c...

Page 4: ...up button Heater Control Ultrasonic lights will display green Operating temperature is pre set 150 F on the TEMP CONTROL Initial heat up time using cold water is approximately two hours For automatic...

Page 5: ...UltraSonic LLC Model Ultra 6000FLT 5 Operation Instructions Control Panel...

Page 6: ...tart Stop Agitation Timer 6 Filtration Manual Start Stop Filtration Timer 7 Auxiliary Manual Start Stop Timer 8 Up Agitation Table 9 Down Agitation Table 10 Menu Machine Timers and Settings See Progra...

Page 7: ...ON Heat Button Manual Start Stop Heating Ultrasonic Button Manual Start Stop Ultrasonic Operation Agitation Button Manual Start Stop Agitation Operation Filtration Button Manual Start Stop filtration...

Page 8: ...button until desired Menu To return back to a Menu selection press down button to return To select desired Menu Option press Menu Button until edit position is reached then use Up and Down Buttons to...

Page 9: ...Button to move to next setting Day AUX Control Set In Menu Press UP Button until Day AUX Control press Menu Button until Day Time On Off use Up Down Buttons to set ON or OFF Press Menu Button to reach...

Page 10: ...in tank is slowed Most users regularly replace their filter each month to avoid any decrease of flow Solution should be replaced regularly to avoid a drop in cleaning performance that will result from...

Page 11: ...ed earlier for refilling and start up procedures MAXIMIZING ULTRASONIC PERFORMANCE Temperature Temperature has a dramatic effect on the cleaning performance and time needed to clean a part The recomme...

Page 12: ...ntaminated Solution As contaminants build up in the solution the ultrasonic performance drops The cavitation bubbles waste their energy imploding on the contaminants and not on the part surface It is...

Page 13: ...LING ULTRASONIC LLC 513 502 9746 SPECIAL NOTE PLEASE RETAIN ORIGINAL SHIPPING CRATE Our service tech will help troubleshoot any problem and instruct user on the safe and proper procedures for repairs...

Page 14: ...n no case shall UltraSonic LLC be responsible for any incidental or consequential damages This warranty does not apply to Equipment or components subjected to misuse improper installation alteration m...

Page 15: ...d _________ Voltage Supply to Machine with heaters on L1 to L2 ___________ L2 to L3 ___________ L1 to L3 ___________ L1 to Ground _________ L2 to Ground ________ L2 to Ground _________ I have received...

Reviews: