The cabinet power supply can be provided using an H05-RNF-3G0.75 type cable.
It is recommended to install a disconnectable bipolar circuit breaker interrupter with visible cut-off of 10 A
upstream of the flow meter power supply. It is also recommended to install lightning arresters on each of
the inputs/outputs of the flow meter, along with a galvanic isolator.
Important:
The wiring and unwiring of the flow meter must be performed by a person with electrical
accreditation. This accreditation must be obtained before dismounting the plate including the keypad
and the display. It is essential for this plate to be closed with a tool after each intervention.
Connection and disconnection from the outlets must be carried out with the power off and the
equipment isolated.
Ultraflux accepts no responsibility for incidents which may occur following a failure to respect these
instructions.
Comments:
Check the marking on the cabinet to ensure its supply voltage.
The flow meter is intended for an overvoltage category II installation.
1.11 Environment for use of the cabinet
Cabinet: –25°C to 50°C
Hygrometry: 80% maximum
Ventilation: no specific precautions.
Warning:
This flow meter is a class A device. In a residential environment, it may cause radio-frequency
interference. In this case, the user may be asked to take appropriate measures.
1.12 Cleaning the flow meter
Use a dry cloth to clean the screen and the keypad.
The box may be cleaned with water, diluted alcohol or detergent using a sponge or a soft cloth. Do not use
abrasive materials or solvents.
1.13 Composition of a measuring point
Apart from the electronic converter (the cabinet), a measuring point includes the following elements:
1 to 8 pairs of probes
1 to 16 special cables for the probes (provided by Ultraflux)
1 to 4 level measurements per channel (which may or may not be provided by Ultraflux)
Summary of Contents for UF 831
Page 5: ...5 NT UF 831 CO GB b March 2014 CHAPTER 1 OVERVIEW...
Page 10: ...10 NT UF 831 CO GB b March 2014 CHAPTER 2 SAFETY INSTRUCTIONS...
Page 14: ...14 NT UF 831 CO GB b March 2014 CHAPTER 3 INSTALLATION AND WIRING...
Page 22: ...22 NT UF 831 CO GB b March 2014 CHAPTER 4 IMPLEMENTING A MEASURING POINT...
Page 26: ...26 NT UF 831 CO GB b March 2014 CHAPTER 5 USING AND CONFIGURING THE UF 831...
Page 59: ...59 NT UF 831 CO GB b March 2014 CHAPTER 6 FUNCTION ENGINE...
Page 61: ...61 NT UF 831 CO GB b March 2014 CHAPTER 7 PC SOFTWARE...
Page 77: ...77 NT UF 831 CO GB b March 2014 Appendix 1 Characteristic echo signals...
Page 80: ...80 NT UF 831 CO GB b March 2014 Appendix 2 Speed of sound in water...