NT214B GB1
20 / 26
8-11 : Optional menus or readings concerning MiniSonic P-B two-pipes version
The optional menus concern the second pipe and permit to validate the second probes
connector. The first and more important is in the Echo menu : Nb of Chords = 1 or 2
The MiniSonic P-B can calculate a Q
T
result from Q1 & Q2 : sum , difference … Its
calculation formula must be entered by using our PC Software .
Concerning specific menus or displays available on LCD and Keypad , we can mention :
Measurement Display menu :
- Display of distance D.S. 1 & 2 for the two sets of probes (same reference)
- Display of calculated flow Q
T
from entered formula
- Display of Gains 1 & 2 plus I.Q. 1 & 2 for the two pipes.
- Display of the two different flows Q1 & Q2 in the two pipes.
- Display of the two flow velocity and also the two sound speeds.
- Display of the two time of flight and the two Delta T on the two pipes.
- Display of the two hydraulic information’s Kh 1 & 2 and Reynolds number Rey 1 & 2
Calibration Allowed menu :
identical to MiniSonic-P version
Pipe Settings menu :
- Entering for pipe 1 & 2 diameter and circumference: ext D 1 = xxxx.x / ext D 2 = xxxx.x
- Entering of pipe 1 & 2 materials: Pipe 1 = / Pipe 2 = …
- Entering of pipe 1 & 2 thickness: Thickness 1 = xx.x / Thickness 2 = xx.x (mm)
Q Settings menu :
identical to MiniSonic-P version . These data are common to the two
pipes.
Probe / Echo Settings menu :
- Number of pipes / chords = 1 or 2
•
If transducer choice is SE, mounting mode ( / , V , N , W ) must be the same for
the two pipes
•
If choice is SM, it is possible to set separately the distances L 1 & 2 and D & & 2
- Management of amplification gains 1 & 2: Gain 1 = ESC … / Gain 2 = Auto …Manual
- Management of the two gains in Manual mode: Gain 2 = xx dB or
- Management of the two margins in Auto mode: Margin 2 = xx dB
General Settings menu :
identical to MiniSonic-P menu but
- Possibility to copy Q
T
or Q1 or Q2 as a 4 – 20 mA output
- Possibility to set Tot 1 or Tot 2 on flows Q& or Q2 or QT
Zero / Auto-Zero Settings menu:
- Possibility to do the Auto-Zero separately on pipes 1 & 2
- Possibility to set or reset separately Delta To 1 & 2
Ouuput current Settings , Factory settings and I/O Test menus:
as MiniSonic-P
Echo Display menu:
it put multiplexing off for signal controls with oscilloscope.