5 – CONVERTER FIRST HANDLING INSTRUCTIONS:
®
Charging Batteries - Charger/ A Auxiliary supply
Cords : EU – Optionally other ( US …)
Approval = CE and CB/UL
AC Input Voltage : 100 -240 V 47/63Hz
Output Voltage : 15 V DC - 1 A
Charging time = 14 hours
A green LED is On when supplied.
Batteries : NIMH – No memory effect
Autonomy: up to 35 h. Depends on use
( 4-20 mA ; backlight … )
MiniSonic–P has a battery level
indicator (%). It takes its calibration
from a complete cycle 100 % down to
0% (Real stop occurs on V Batt < 8V)
Be sure that batteries are sufficiently charged before you start your operations.
®
Please read § 7 and § 8 before any measurement try.
®
Select one probes set depending on your application (see § 6)
®
Start settings and controls in following order:
- Switch the MiniSonic-P on and wait for the welcome displays.
- Enter in “Probes / Echo menu” and set correct transducers with mounting mode
- Enter in “Pipe settings “ menu and enter corresponding data .
- Enter in “Q settings“ menu and confirm liquid characteristics .
- After the measurement menu has displayed requested distance between probes,
- Fix these probes with coupling gel or tape and do connections to MiniSonic - See § 6
- Read results and attached information in “Measurement menu” - See § 8-1
- If not satisfying, proceed to possible corrections - See § 9
- Start measurement records or hand note results upon your own choice.