8-6 : « General settings & Relays / Outputs » menu :
∇
Setting of LCD backlighting:
(Depending on MiniSonic version)
∇
Filter rate of measurements
(Number of 0.5 s cycles)
∇
Last accepted measure.
Memorising time (if IQ < 33%) =
(mode ESC : enter Mém. = > 20)
∇
Setting of Relays 1 & 2
Alarm C +/- set two thresholds on C
Relays switch
On
at set point and
Release after an Hysteresis of 2%
! Current outputs are available only
on MiniSonic-ISD
∇
Settings of current output 1
Current output is reserved to the
Interface detection information
S
et in relation with density range
∇
Settings of current output 2
Current output 2 can copy the
Ultrasonic signal result ( Gain=Auto)
It gives information on mixtures as
it flows in a drainage tube .
∇
RS 232 / 485 Jbus Slave Nbr
By Default, our Software asks 1
(Possibility up to 255
– Zero= none)
∇
RS 232 / 485 Baud Rate
(300, 600, 1200, 2400, 4800)
Rugosité (mm)
*.**
Relay 1 = Sphere
(closed , open , fault , alarm…)
Filter =
∗∗
Mémory (s)=
∗∗
Back Light
ON / OFF / TIMED
Relay 2 = Fault
(sphere , alarm
– C , alarm +C )
O.C.1
C 4 mA = xxxx.x m/s
O.C.1
C 20 mA = xxxx.x m/s
O.C.2
Gain 4 mA = xx dB
O.C.2
Gain 20 mA = XX dB
N. JBUS = 1
BAUD = 9600