NT216C GB1
18 / 26
8-2
: « Calibration Allowed »
Menu :
> Access = press
F
once from the Measurement Display menu
If the display is:
Enter this code (1 to 65535)
If the display does not request “Code?”, there is none (code = 0 )
> Press
∇
Enter a code:
! This code will become active after de-energizing.
Remember to make a note of it. It will be required for any subsequent
intervention.
> In case of loss, contact Ultraflux giving the following informations; the serial number of
your MiniSonic MK4*/xx/xx/xxxx as it appears during energizing and the date or dates
on which you want action to take place. An operation will be requested allowing a
provisional code to be calculated.
> A second press on
∇
will provide access to the existing Languages choice field.
8-3 : « Probe / Echo Settings » menu :
> Access = Press twice on
F
or
F
once, then
+
and Enter by
∇
(operations common to all the menus).
∇
Probe Frequency (MHz)
(Choice = O.25 , 0.5 , 1 or 2 )
∇
Dead Time To through probe face
(Typical value = 4 to 6 microsec. )
It includes time in coupling tape
∇
Amplification Gain mode
We recommend ESC mode
Other options :
Auto
and Manual
∇
If Auto : enter over amplification
Recommended value = 12 to 16 dB
∇
If Manual : enter Gain in dB
Freq. = 1 MHz
To = xx.x Micro Sec
Gain = ESC
( Auto / Manual )
Margin = xx dB
Calibration Allowed : 0
Code ?
Access Code = xxxx
Gain = xx dB