FACILITY & ENVIRONMENTAL CONSIDERATIONS
GUIDELINES
In addition to receiving proper training in the use of soft-tissue dental lasers, users should be familiar and experienced with
these procedures using electrosurgical devices or traditional instruments before performing them on patients with the Gemini
EVO 810+980 soft tissue laser. Non-experienced users should seek appropriate training guidance before attempting clinical
treatments with the Laser system.
In order to insure the safe use of the Gemini EVO 810+980 soft tissue laser in your facility, please check to make sure that the
proposed location is compatible with the specifications listed below.
POWER REQUIREMENTS
External AC/DC Power Supply - Use only the provided Gemini EVO laser power supply. Every Gemini EVO laser power supply
shows the corresponding label below. DO NOT use any other power supply.
Input Power: 100-240V; 50-60Hz, 1.5A
Output Power: 18V, 65W
HEATING AND VENTILATION
Operating environmental conditions to be within 10° - 40°C (50° - 104°F), and 95% relative humidity or less. Transportation
and storage environmental conditions to be within 0° - 40°C (32° - 104°F), and relative humidity of 95% or less. Atmospheric
pressure to be within 70kPa – 106kPa in operating, transportation and storage conditions.
COMBUSTIBLE CHEMICALS AND GASES
All gases that are combustible or support combustion and are used in the operatory area where the Gemini EVO 810+980
soft tissue laser is being operated must be turned off during the procedure. Cleaning supplies or other flammable chemical
compounds should be stored in an area away from the surgical site in order to avoid possible combustion. Do not use in the
presence of supplemental therapeutic oxygen supplies for patients with respiratory or related diseases.
PLUME EVACUATION
Plume evacuation should be addressed when vaporizing tissues. A high volume vacuum system should be used and 0.1
micron or less high filtration masks that are suitable for virus and bacterial control should be worn by clinicians.
OPERATORY ACCESS DURING LASER USE
Access to the treatment area should be restricted while the lasers are in use. A sign indicating “LASER IN USE” should
be placed in a designated area adjacent to the treatment area entry location.
PLEASE USE THIS POWER
SUPPLY
ONLY
WITH
GEMINI
EVO
LASER
PLEASE USE THIS POWER
SUPPLY
ONLY
WITH
GEMINI
EVO
LASER
G2-9-019 - REV 1.0
CAUTION:
LASER FUME AND/OR PLUME MAY CONTAIN VIABLE
TISSUE PARTICULATES