Control Solution Results
The iPet PRO meter is functioning
correctly only if the control
solution test result is within the
specified range printed on the test
strip vial. If the test result is out of
range, please repeat the test.
Out of range result may be
caused by:
•
Incorrect steps of performing
the test
•
The control solution temperature
is lower than 59°F (15°C) or
higher than 95°F (35°C)
•
Expired or contaminated
control solution
•
Expired or contaminated
test strips
•
Meter malfunction
Note:
•
The result will not be included
in day average calculation when
the iPet PRO meter is set in the
control solution test
mode.
•
DO NOT use iPet PRO meter
if the problem persists. Please
contact customer service
immediately.
Memory Features
How to view results stored
in memory
1. Press to turn on the meter and
press again to enter the animal
type selection mode. Press or
to select a desired animal type
(Fig.38).
2. After selecting a desired animal
type, press to confirm.
3. Press to view previous results.
The “MEM” icon will be displayed
at the top. Initially, the “year”
will be displayed (Fig.39). After 1
second the month, day, and time
will then be displayed (Fig.40).
The most recent memory results
will be displayed first.
4. You can press or to scroll
forwards and backwards through
the results.
5. Press button to exit the stored
test results and view the control
solution test result (Fig.41).
6. When the meter displays the last
stored test result, press button
so the meter will display day
average results (Fig.42).
Note:
The meter stores 500 results in the
memory. When the memory
is full, the oldest result will be
removed and replaced by the
most recent result.
(Fig.38)
(Fig.39)
(Fig.40)
(Fig.41)
(Fig.42)