UIM-90 (221)
UIM-90 (221/В)
UIM-90 (221/Р)
UIM-90 (221/ВР)
22
c) after the end of the test, the results are automatically printed out on
the built-in thermal printer
*
(if the parameter "Printing" is set to ON), and the
LCD display will show:
,
after which the LCD display will show the test results:
1) the parameter "Number of break. N =" is
set equal to one;
2) the parameter "Number of break. N =" is
set greater than one;
d) repeat steps (f) to (h) of paragraph 4.2 to view the test results;
e) open the test compartment lid, remove the cell;
f) to conduct the next test, prepare the sample in accordance with
paragraph 4.1 and repeat steps (b) to (e) of paragraph 4.3.
If the tests are complete, close the test compartment lid. Set power
switch
in position "О", green LED indicator
will turn off. Unplug power
supply cable from the network socket (in case of mains operation).
*
- for models UIM-90 (221/P); UIM-90 (221/ВР)