How to install the Flange Assembly:
1.
Use an appropriate CONFLAT Gasket
2.
Use 6 appropriate length Mounting Bolts and washers
(note: only hex head bolts can be
used)
3.
Place CONFLAT Gasket into flange counter bore
4.
Insert MD40 flange assembly
5.
Insert bolts and tighten.
Note: when tightening bolts, care must be taken not to tighten them
unevenly as this may cause the gasket not to seal properly.
STAGE 3
| Re-fitting the Drive Thimble and Motor Assembly
Refitting the Drive Thimble
Actuator Housing Screws
3 x M3 Grub Screws
Bearing Cap
Drive Thimble