Page 11
37215 Instruction Manual (Rev. 4)
6.3
Transforming a standard 37215 into a Mouse model
The optional kit
37215-327
(see paragraph 7.2) includes the necessary components to
transform a standard 37215 into a mouse specific device, with reduced (50%) applied
forces, see also paragraph 4.3-Reading the Results.
The kit includes:
A specific slide
A chisel-shaped pusher
Two 50% reduced additional weights
The replacement is easily done, following the steps indicated in the drawing below:
1)
slacken the two screws
and the remove the
plastic stopper
2)
Pull-out the standard
slide and position the
mouse one.
3)
Replace the standard
pusher with the chisel-
shaped one (flat)
4)
Use the reduced addi-
tional weight if needed
If an Analgesy-Meter with digital reading is in-
volved, replace the standard pusher tip with the
flat tip
standard
flat