Page 6
37215 Instruction Manual (Rev. 4)
3.4
Connections
Three low voltage connectors (for maxi-
mum safety of the operator) are located
on the right side of the back panel of the
instrument, which also lodges the mains
module (see drawing 2 in paragraph
3.3), labelled:
START-STOP SIGNAL
:
the start-stop signal connection enables the operator, to connect
the Analgesy-Meter to the optional Analgesy-Meter DAQ 37215-
100, see paragraph 5-DATA ACQUISITION and 7.2-Optional (
*
)
PEDAL SWITCH:
insert the male connector of the pedal switch cable until it locks into
its socket; a “click” confirms it is locked;
GEARED MOTOR:
this connector supplies low voltage power to the geared motor. In-
sert the male connector of the geared motor until it locks into its
socket; a “click” confirms it is locked.
Accidental mismatch does not cause any circuit damage.
(
*
) For connection to the optional Analgesy-Meter DAQ 37215-100, please refer to the re-
lated instruction manual.
3.5
Intended Use
The Analgesy-Meter is intended for investigation use on laboratory animals only.
3.6
General Safety Instructions
The following guidelines must be followed to ensure safe operation.
DO NOT
attempt to open or perform any service work
DO NOT
connect up human subjects
3.7
Additional Safety Consideration
a.
Use original accessories and spare parts only, see also paragraph 7-
ORDERING INFORMATION.
b.
immediately disconnect and replace damaged mains cord.
c.
do not operate in hazardous environments or outside prescribed environmen-
tal limitations (i.e. +10c° / +40c°, 95% max. relative humidity, non-condensing),
see also paragraph 7.2-Optional.
d.
do not spray any liquid on the connectors and on the geared motor, see also
paragraph 6-MAINTENANCE.
UGO BASILE DOES NOT ACCEPT ANY RESPONSIBILITY FOR PROBLEMS OR
HARM CAUSED TO THINGS OR PERSONS, ARISING FROM: