Instructions for use (rev 05/18)
e.ON SINTER
UGIN DENTAIRE
43
8.1.
Heating element replacement
Observe these conditions for replacing heating elements:
-
preferably, be at room temperature (+15°C / +25°C), the heating elements are very sensitive to
temperature;
-
handle them carefully;
-
keep all elements of heating elements: heating elements and cushioning foam pad protecting.
Steps to heating element replacement:
1.
Open the furnace upper cover
(unscrew)
2.
Unscrew the electrical contacts of heating
elements with a wrench (supplied as an
accessory)
3.
Repeat step 2 for all the heating elements
to replace
4.
Unscrew the support elements of heating
element with a wrench (supplied as an
accessory)
5.
Remove carefully the
cushioning foam pad
protecting heating elements.
Lift carefully the heating
element with the elements
protecting.