![Ugin e.on sinter base Instructions For Use Manual Download Page 20](http://html1.mh-extra.com/html/ugin/e-on-sinter-base/e-on-sinter-base_instructions-for-use-manual_3247399020.webp)
Instructions for use (rev 251121)
e.ON sinter base
UGIN DENTAIRE
19
ELEMENTS - CHAMBER
(see chapter 6.7.1
–
6.7.2)
-
Place the sole (2 units) and insulating plate
while the furnace is working.
-
Observe the information about the positioning
of elements
(chapter 6.7).
-
Always place the elements in the tray for
sinterizing, with the zirconia beads inside, and
place the cover
.
-
Beads for sintering must be changed in any tray
if a
discoloration in yellow colour
appears.
DECONTAMINATION PROGRAM
(see chapter 6.6)
-
Before first sintering, user must launch a
decontamination program (No. 40).
-
Always place the sole (2 units) and the
insulating plate in the chamber during carry
out a decontamination program.
-
Add in a tray with or without a cover white
Zirconia smashed in multiple small pieces
(absorption of pollutants) or a powder specially
designed to improve the effectiveness of
decontamination.
After running the program,
check the condition of the rods:
-
still whitish traces = still contaminated:
repeat a decontamination cycle;
-
rods with a grey shiny appearance = ok:
furnace ready.
-
High translucency zirconia
are very sensitives
(crossed discoloration risk): user must do
regularly decontamination program.
-
If
liquid shades
are used with staining technic,
prosthetics units must be dry and preheated
with an infra-red lamp or into an oven to
prevent pollution into the
eON sinter base
, in
following zirconia and shades manufacturer
recommendations.
-
White clusters can appear on the rods
(heating
elements) when different brands or types of
zirconia are used: user must clean chamber and
rods with a special zirconia cleaner like
Nacera
Clean (registered brand)
with decontamination
program.
Note: observe the instructions set out in chapter 8
–
maintenance.