background image

15

b

AttEriEs

 r

EpLACEmEnt

 

This meter has been designed for use with both alkaline 

 

and rechargeable Nickel Metal Hydride (NiMH) batteries. 

 

No other types are recommended. The analyzer is supplied  

 

with 4 “AA” size alkaline batteries. 

 

These should be installed into the instrument as shown  

 

in the diagram to the right and indicated on the back of  

 

the unit.

 

CAUTION!

 

 

Take great care when installing the batteries to observe  

 

correct polarity. Always check the meter for operation 

 

immediately after installing new batteries.

U

sinG

 r

E

-C

hArGEAbLE

 b

AttEriEs

 

 

The battery charger must only be used when NiMH batteries are fitted. Alkaline batteries 

are not  

 

re-chargeable. Attempting to recharge alkaline batteries may result in damage to the product and may  

 

create a fire risk.

b

AttEry

 C

hArGinG

 

Ensure that you use the correct charger. This unit uses a 9V DC regulated charger.

 

Ensure that the batteries are fitted in the correct manner, and then charge for at least 16 hours. Subsequent  

 

charges should be overnight. NiMH batteries may be charged at any time, even for short periods to  

 

conduct testing.

 

WARNING!

 

 

Under NO circumstance should you expose batteries to extreme heat or fire as they may explode and cause 

 

injury. Always dispose of old batteries promptly in a manner consistent with local disposal regulations.

E

LECtrOmAGnEtiC

 C

OmpAtibiLity

 (EmC)

 

This product has been tested for compliance with the following generic standards: 

 

EN 50081-1, EN 50082-1 and is certified to be compliant.

 

The European Council Directive 89/336/EEC requires that electronic equipment does not generate  

 

electromagnetic disturbances that exceed defined levels and has an adequate level of immunity to enable it to 

 

be operated as intended.

 

Since there are many electrical products in use that pre-date this Directive and may emit electromagnetic  

 

radiation in excess of the standards defined in the Directive there may be occasions where it would be  

 

appropriate to check the analyzer prior to use. The following procedure should be adopted.

     

  •  Go through the normal start up sequence in the location where the equipment is to be used 

 

    •  Switch on all localized electrical equipment that might be capable of causing interference

 

  •  Check that all readings are as expected (a level of disturbance in the readings is acceptable)

 

    •  If not, adjust the position of the instrument to minimize interference or switch off, if possible, the  

 

     offending equipment for the duration of the test

 

At the time of writing this manual (July 2009) UEi is not aware of any field based situation where such  

 

  interference has ever occurred and this advice is only given to satisfy the requirements of the Directive.

Note: Follow battery 

directions on back of 

housing

Battery Compartment

Cover

Back of

Instrument

C75 Directions

C125/C127, C155/C157 

Directions

OR

Summary of Contents for EagleX

Page 1: ...1 Eagle Eagle X Owners Manual...

Page 2: ...Charge Indicator Light 4 Line Backlit Display Display Line Lights Particle Filter Inside Water Trap Water Trap Protective Rubber Boot w built in Magnets Flue Probe Temperature Plug Plugs into T1 Narr...

Page 3: ...ulated heat lost turning the Carbon in the fuel to Carbon Dioxide CO2 Wet Calculated heat lost turning the Hydrogen in the fuel into water H2O CO Loss Calculated loss due to partially burnt Carbon Any...

Page 4: ...as soon as possible Use the GAS ZERO function Eagle X only to purge the analyzer of excess gases To do this on a standard Eagle remove the probe from the flue and turn ON the pump Always allow the rea...

Page 5: ...ng testing Connect flue probe thermocouple connector to T1 and connect flue probe to water trap as shown above Use optional probe with T2 for inlet temperature Insert Flue Probe in stack Adjust the co...

Page 6: ...raft probe for measuring pressure Press and hold t DOWN button to zero pressure sensor Rotate test selector to Temp Connect flue probe thermocouple or accessory thermocouple connector to T1 Connect ac...

Page 7: ...ss Air O2 and CO readings both before and after the blower turns on If the heat exchanger is sealed your O2 and CO readings should remain fairly stable A breach in the heat exchanger may allow fresh a...

Page 8: ...inch mmHg Millimeters of mercury hPa Hecto pascals EXIT Return to main menu REPORT COMB N VIEW When viewing reports PRESSURE DEL ALL Press s UP or t DOWN to select report EXCH Press and hold s UP or t...

Page 9: ...04 to 06wc Overfire Draft Water Column Inches 02wc Carbon Monoxide parts per million 100ppm Smoke 0 or manufacturer s recommendation Positive Overfire Gas Oil Oxygen 3 to 9 O2 Stack Draft Water Colum...

Page 10: ...Verify proper combustion O2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Send and Return Water temp 80 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas...

Page 11: ...mperature Rise AC side Static Pressure Drop across coils Oil Furnace Verify proper combustion O2 CO Stack Temp Stack Draft Efficiency Test Verify Smoke Set Up Over Fire Draft Verify proper Static Duct...

Page 12: ...ce between the ambient pressure level and the pressure level in the flue This is created either by the natural buoyancy of the hot gases created in combustion lifting or by an inducer fan that assists...

Page 13: ...product of combustion and from wet fuel is recovered and the wet loss term is zero Gross efficiency calculations assume that the energy contained in the water vapor is not recovered Since the fuel air...

Page 14: ...vents or strong detergents as they may damage the finish impair safety or effect the reliability of the structural components Emptying Cleaning the In Line Water Trap The in line water trap should be...

Page 15: ...ations Electromagnetic Compatibility EMC This product has been tested for compliance with the following generic standards EN 50081 1 EN 50082 1 and is certified to be compliant The European Council Di...

Page 16: ...rights You may also have other rights which vary from state to state U S A 1 8 0 0 5 4 7 5 7 4 0 F a x 5 0 3 6 4 3 6 3 2 2 C A N A D A 1 8 7 7 4 7 5 0 6 4 8 F a x 6 0 4 2 7 8 8 2 9 9 W W W U E i T E S...

Reviews: