4
s
AfEty
n
OtEs
Before using this meter, read all safety information carefully.
“WARNING” is used to indicate conditions or actions that may pose physical hazards to the user.
“CAUTION” is used to indicate conditions or actions that may damage this instrument.
WARNING!
This analyzer extracts combustion gases that may be toxic in relatively low concentrations. These gases are
exhausted from the back of the instrument. This instrument must only be used in well-ventilated locations.
It must only be used by trained and competent persons after due consideration of all the potential hazards.
p
rEfLiGht
C
hECkList
• Clean particle filter
• Water trap and probe line are empty of water
• Power on and zero
• All hose and thermocouple connections are properly secured
• Flue gas probe is sampling ambient FRESH air
• Water trap is fitted correctly to the instrument
• Flue temperature plug is connected
• Inlet temperature probe is connected if required
A
nALyzEr
C
OnnECtiOns
NOTE: Take care when inserting the temperature probes as the pins are polarized. Insert with the smaller
pin (+) to the right.
WARNING!
Turning the pump off while the probe is in the flue will leave toxic gases inside the analyzer. Once data
has been printed or copied it is advisable to purge the unit with fresh air as soon as possible. Use the GAS
ZERO function (Eagle X only) to purge the analyzer of excess gases. To do this on a standard Eagle remove
the probe from the flue and turn ON the pump. Always allow the readings to return to zero (20.9 for O2)
prior to shutting the unit off. The meter will not switch off until the CO reading is below 20 ppm.
WARNING!
The probe will be hot from flue gases. Remove the probe from the flue and allow it to cool naturally. Do
not immerse the probe in water, as this will be drawn into the analyzer and damage the pump and sensors.
Once the probe is removed from the flue and the readings have returned to ambient levels hold down
“On/Off” and switch off the analyzer. The instrument will count down from 30 to switch off. If
you pressed “On/Off“ by mistake, pressing “Send“ will return you to normal operation.
Getting Started
• Setting Inlet Temperature
- Turn on and zero the analyzer with
out the flue probe connected to use
ambient temperature
- Connect flue probe thermocouple to
T1 during zero countdown to store
probe tip temperature as inlet
(ducted system)
p
Ost
f
LiGht
(1)
Remove the probe from the flue and allow the analyzer to purge with fresh air until readings return to zero.
- O2 to 20.9%, CO to Zero
(Be careful as the probe tip will be HOT)
(2)
Drain water trap
(3)
Check particle filter
Drain water trap by unplug-
ging the drain plug and
shake to get excess water
out.
Check particle filter for dirt
and any other sediment and
replace if necessary.