8
L. Test Lead Input Jacks:
Multifunction and Positive input jacks.
• Multifunction input port used for measuring: AC or DC volts, resistance,
continuity, diode, capacitance, AC or DC µA.
M. Wire Separation Tab/ NCV sensor:
Use to isolate an individual wire from
a bundle for testing. NCV sensor detects live voltage.
N. Test Lead Holder
O. Hand Guide:
Used as a point of reference for the operator’s safety.
P. Hold/Worklight/ Back light Button:
• Press to hold the reading on the display. Press again to return to live
reading.
• Press and hold to turn on Worklight and Back light.
• Worklight and Backlight turn off after 60 seconds.
• Press hold during power up to disable the Auto power off function.
Q. Display:
• High contrast dual display with backlit.
• Amps (AC/DC) reading will always display on upper display.
R. LRA Inrush/Zero/Hz/Duty Button:
• Press to Zero the DC Amps reading.
• Press to enter LRA Inrush mode (See page 17 for details)
(must be in AC Amps mode first).
• Press again to return to live readings.
• Press and hold to select Hz/Duty in AC volt or AC μA.
S. K-Type Temperature Probe Inputs:
T1 (Left) and T2 (Right))
T. Input Jack Lock:
Switch to use Temperature or Test lead inputs
U. Built-in versatile magnet to use as a mount or as a strap
V. Battery Cover:
Easy access for replacing batteries without breaking
calibration seal.
W. Battery Compartment Latches:
X. Serial Number
Overview (Cont.)
Low Z (Low Impedance)
• Rotate Selector Dial to Low Z
• Default = Auto Selection (AC or DC)
• Press SELECT x1 =
• Press SELECT x2 =
Features:
BT
SELECT
30V
T
1
TEMP
T
2
True RMS
DL589
PEAK
µA
Temp
600V
600A
CAT IV
V
A
SELECT
Low Z
AUTO
V
A
Low Z