AC/DC Microamps: <2000µA
WARNING
• Do not attempt to measure more than 2000µA.
Features:
BT
SELECT
DC Microamps Measurement -Test lead input
Range
Resolution
Accuracy
Overload Protection
600µA
0.1µA
±1.2% + 3dgts
600V RMS
2000µA
1µA
AC Microamps Measurement -Test lead input
Range
Resolution
Accuracy
Overload Protection
600.0µA
0.1µA
±1.5% + 3dgts
600V RMS
2000µA
1µA
45Hz to 400Hz True RMS
Temperature Calibration
The new DL589 clamp meter offers a digital field temperature calibration
procedure to make the process easier. Here are the steps.
1) Set dial position to “°F°C” and check T1 mode
2) Press and hold SELECT button for about 2 seconds
3) Press HOLD button while “FCAL” is displayed on top of the LCD to enter
Field Calibration.
4) After immersing T1 temperature probe in ice bath and the temperature is
stable, press and hold Hz/Duty button
5) Press SELECT button to switch to T2 mode.
6) After immersing T2 temperature probe in ice bath and temperature is
stable, press and hold Hz/Duty button to complete Field Calibration