background image

 WARNING

•  Do not attempt to measure more than 2000µA DC

Features:  

  

  

DC Low Amps – Test Lead input

Range

Resolution

Accuracy

Overload Protection

600µA

0.1µA

±(1.2% +3 dgts)   

2000µA/600V RMS

2000µA

1µA

Temperature C˚/F˚

DL479

NCV

°C

°F

CAT IV 600V

CAT III 1000V

MAX

µA

A

 CAT Ill

1000V                     600A

True RMS

Hz / Duty

MAX/MIN

 Disconnect test lead probes from voltage source and meter.

•  Press 

Select button

 to change between Fahrenheit and Celsius

Features:   

  

Range

Resolution

Accuracy

Overload Protection

-31° to 752°F

0.1°F

±(1.5% +3.6°F )

600V RMS

-35° to 400°C

0.1°C

±(1.5% +2 °C)

Sensor must be thermocouple type.

Stated accuracy does not account for thermocouple accuracy.

•  Default   =  
•  Press x1 =  

Non-Contact Voltage

DL479

NCV

°C

°F

CAT IV 600V

CAT III 1000V

MAX

µA

A

 CAT Ill

1000V                     600A

True RMS

Hz / Duty

MAX/MIN

•  Select NCV mode and move the tip of the clamp meter near voltage source.

  Both an Audible and Visual alert will indicate voltage.
•  Non-Contact Voltage Detection is used to detect power with sensor located 

in the tip of the clamp head indicates positive response with both an Audible 

and Visual alert.

•  Do not use non-contact voltage detector to determine if there is current on 

the wire. Detection operation could be affected by socket design, insulation 

thickness, type or other factors.

•  Voltage indicator light may also light when voltage is present on the meter’s 

input jack or from an external interference such as motors, flashlights, etc.

NCV Sensor

located in tip

On Voltage

Approx. 25V AC

Frequency (Hz) / Duty Cycle

DL479

NCV

°C

°F

CAT IV 600V

CAT III 1000V

MAX

µA

A

 CAT Ill

1000V                     600A

True RMS

Hz / Duty

MAX/MIN

 • Use CAT III test leads or higher.

 •  Set Function dial to AC/DC Amps, select AC Amps, press and hold the 

 

Select button

 for Frequency and Duty Cycle modes.

 •  WARNING

 • Do not attempt to measure more than 750V AC/600V DC.

Features:  

Frequency

Range

Resolution

Accuracy

Overload Protection

99.99Hz

0.01Hz

±(0.1% +3 dgts)

600V RMS

999.9Hz

0.1Hz

9.999kHz

0.001kHz

Duty Cycle

Range

Resolution

Accuracy

Overload Protection

0.5% to 95% 

60Hz to 400Hz

0.1%

± (0.2% per kHz 

+2.0% +2 dgts)

600V RMS

15% to 85% 

400Hz to 2kHz

Frequency Width: 60Hz to 400Hz: 0.5% to 95%

 

400Hz to 2kHz: 15% to 85%

Sensitivity: >6Vpp RMS

• Default =  
• Press and hold  =  
• Press  =  

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for DL479

Page 1: ...ress warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of original purchase date will be required before warranty repairs will be rendered Instruments out of warranty will be repaired when repairable for a ser...

Page 2: ...itance Always adhere to national and local safety codes Use proper personal protective equipment PPE to prevent shock and arc blast injury where hazardous live conductors are exposed Always turn off power to a circuit or assembly under test before cutting unsoldering or breaking the current path Even small amounts of current can be dangerous Always disconnect the live test lead before disconnectin...

Page 3: ... on LCD only during auto ranging mode O Apo Auto power off after 30 minutes of use Press and hold the SELECT button while turning the meter on to disable Apo auto power off P Display High contrast display Q Category Max Indicator Maximum CAT Rating for input jacks R Input Jacks Multifunction and Common jacks Multifunction input port used for measuring AC or DC volts Resistance Continuity Diode Cap...

Page 4: ...load Protection 600mV 0 1mV 1 0 3 dgts 1000V RMS 6V 1mV 60V 10mV 600V 100mV 750V 1 0V True RMS Frequency Range Square 50Hz to 100Hz Sine 50Hz to 400Hz Bandwidth Sine 0 5 error at 1 5kHz max Square 0 5 error at 0 1kHz max Triangle 0 5 error at 1 2kHz max DL479 NCV C F CAT IV 600V CAT III 1000V MAX µA A CAT Ill 1000V 600A True RMS Hz Duty MAX MIN Range Resolution Accuracy Overload Protection 600Ω 0 ...

Page 5: ...9 NCV CAT III 1000V MAX µA A CAT Ill 1000V 400A True RMS AC Amps 600A DL479 NCV C F CAT IV 600V CAT III 1000V MAX µA A CAT Ill 1000V 600A True RMS Hz Duty MAX MIN DL469 NCV CAT III 1000V MAX µA A CAT Ill 1000V 400A True RMS Center wire in conductor alignment marks for best accuracy Opposing currents cancel use line splitter when necessary WARNING Keep hands below guard when measuring high current ...

Page 6: ...th both an Audible and Visual alert Do not use non contact voltage detector to determine if there is current on the wire Detection operation could be affected by socket design insulation thickness type or other factors Voltage indicator light may also light when voltage is present on the meter s input jack or from an external interference such as motors flashlights etc NCV Sensor located in tip On...

Reviews: