Battery Replacement

• 

  When the batteries are too low for safe operation, the Low Battery 

indicator will display.

•  Rotate latches until Unlock symbols are aligned with arrows. Remove 

battery cover.

•  Replace the old batteries with 2 new (AAA) batteries.
•  Replace the battery cover
•  Rotate latches until the lock symbols are aligned with arrows.

S/N:

CA
TIV

600V

CA
TIV

600V

OPEN

LOCK

OPEN

LOCK

OPEN

LOCK

OPEN

LOCK

 

 

 

 

 

 

Test Lead Notes

CatIV 600V Measurement Locations

•  Ensure the test lead shield is pressed firmly in place. Failure to use the   

CATIV shield increases arc-flash risk.

CATII 1000V Measurement Locations

•  CATIV shields may be removed for CATII locations. This will allow testing on 

recessed conductors such as standard wall outlets. Take care not to lose 

the shields

4mm

CATIV

600V

CATIV

600V

CATII

1000V

18mm

4mm

CATIV

600V

CATIV

600V

CATII

1000V

18mm

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for DL220

Page 1: ...eipt or other proof of original purchase date will be required before warranty repairs will be rendered Instruments out of warranty will be repaired when repairable for a service charge For more information on warranty and service contact This warranty gives you specific legal rights You may also have other rights which vary from state to state DISPOSAL CAUTION This symbol indicates that equipment...

Page 2: ... the finger guards molded into the probe Use caution when working with voltages above 60V DC or 25V AC RMS Such voltages pose shock hazards To avoid false readings that can lead to electrical shock replace batteries if a low battery indicator appears Unless measuring voltage or current shut off and lockout power before measuring resistance or capacitance Always adhere to national and local safety ...

Page 3: ...e the meter displays the new value Press and hold the A Hold button when meter is in Auto Hold mode to disable Auto power off mode Press and hold again to enable Auto power off mode When the meter is in Auto Hold mode there will be an audible beep while the meter captures a new stable value LRA Inrush Press and hold to enable LRA Inrush mode Press and hold to return to live readings Used for measu...

Page 4: ...D LRA INRUSH AC DC Voltage 600V AC DC WARNING Do not attempt to measure more than 600V AC DC Keep hands below line when measuring high current levels Do not exceed 600V DC or AC RMS on either the common or multifunction test leads as measured from earth ground Select AC or DC Voltage Features A HOLD LRA INRUSH AC Volts Ranges Resolution Accuracy Overload Protection 1V to 199 9V 0 1V 1 5 5 dgts 600...

Page 5: ...Frequency Range sine wave 45Hz to 400Hz Maximum Displayed value AC220A Minimum Displayed value AC 500mA CAT IV 600V VΩ DL220 CAT IV 600V VΩ DL220 A HOLD LRA INRUSH Single Conductor Only LRA Inrush The meter must be in AC Amps mode first to be placed into LRA measurement mode The UEi LRA Inrush is programmed to properly capture the starting current for compressor motors Press and hold A Hold LRA In...

Page 6: ...V C A T I V 6 0 0 V OPEN LOCK OPEN LOCK OPEN LOCK OPEN LOCK Test Lead Notes CatIV 600V Measurement Locations Ensure the test lead shield is pressed firmly in place Failure to use the CATIV shield increases arc flash risk CATII 1000V Measurement Locations CATIV shields may be removed for CATII locations This will allow testing on recessed conductors such as standard wall outlets Take care not to lo...

Reviews: