30
Before the new calibration can be accepted an authorization code is required.
Enter each number of the code 5128.
Once the last digit has been entered the analyzer will store the new settings into its memory, this may take a
few seconds and you will be asked to PLEASE WAIT. Once the memory settings have been checked the
analyzer will display:
Select to either run a final calibration report or not before returning to the main menu.
If in any doubt about the calibration of the analyzer contact a service centre.
10.4.4 Printed Calibration Report
During all stages of the calibration routine you will be asked if you would like to print a calibration report.
See section 6.1 for details on printing.
The calibration report is shown below
ENTER SERVICE CODE
5 1 2 8
CALIBRATION RESET
PRINT REPORT NO
UEi
Calibration Report
Version 2.4
Date: 07-08-01
Time: 09:10:31
Max:
14.7
CO2 % vol:
13.6
Min:
13.3
Max:
3.68
CO % vol:
3.63
Min:
3.32
HC CAL Gas:
2000
PEF:
0.512
Max:
1075
CO2 % vol:
1024
Min:
973
Max:
1050
CO2 % vol:
1006
Min:
950
AMB deg:
24.4
PRS mbar:
1068