Technical information
Technical information
E
E
1. Error messages
Failure encoun-
tered
Possible cause
Corrective action
Display Er 1
Sensor error or faulty
pressure pump
If the fault occurs several
times, contact the customer
service centre.
Display Er 2
Device could not detect
a pulse or could not cal-
culate blood pressure
values.
Check the hose line and air
plug for blockages. Repeat
the measurement. If the error
occurs again, please contact
the customer service centre.
Display Er 3
Deflation rate too high
or too low.
Repeat the measurement. If
the fault occurs several times,
contact the customer service
centre.
Display Er 4
The cuff takes a long
time to inflate. The cuff
is not sitting correctly
or the connection to the
tube is not airtight.
Fasten the cuff correctly and
check that the air plug had
been fully inserted into the
device. Repeat the measure-
ment.
Display Er 5
The air hose is kinked.
Unkink the air hose. Repeat
the measurement.
Display Er 6
The device has detected
extensive pressure
fluctuations.
Repeat the measurement. Do
not move. Do not speak.
Display Er 7
The pressure measured
by the sensor is over the
limit value.
Repeat the measurement. If
the fault occurs several times,
contact the customer service
centre.
Display Er 8
The limit is incorrect
or the device does not
have a limit.
Repeat the measurement. If
the fault occurs several times,
contact the customer service
centre.
No display after
unit is switched on
Batteries inserted
incorrectly.
Check position of batteries.
Batteries flat.
Change batteries.
Err is shown on
the display, device
switches off
Arm movement as the
cuff was being pumped
Repeat the measurement. Do
not move.
Talking during the
measurement
Repeat the measurement. Do
not speak.
2. Customer service
3. Technical data
Device repairs may only be performed by the manufacturer or an ex-
pressly authorised body. Please contact:
UEBE Medical GmbH
Bürgermeister-Kuhn-Straße 22
97900 Külsheim
Germany
[email protected], www.uebe.com
Model:
visomat comfort eco
Size:
L = 110 mm x W = 118 mm x H = 52 mm
Weight:
225.5 g
Display:
LCD display (liquid crystal display) 65 x 48 mm
Memory:
2 memories each capable of storing 90 readings
(stored automatically) and average value (AVG)
Measurement procedure: Oscillometric determination of systole, diastole
and pulse
Pressure display range:
0-299 mmHg
Measurement range:
Blood pressure measurement: 0 - 280 mmHg
Pulse measurement: 40 - 199 pulses/minute
Measurement precision:
Blood pressure measurement (cuff pressure): ± 3
mmHg, Pulse rate: ± 5%
Power supply:
4 x 1.5 V AAA LR03 alkaline manganese batteries
Optional: visomat U2MC mains adapter, output 5
V DC, 1 A
Operating conditions:
Ambient temperature 5 to 40 °C, rel. air humidity
15 to 93 %, air pressure 700 to 1060 hPa
Storage and transport
conditions:
Ambient temperature -25 to 70 °C, rel. air humidi-
ty ≤ 93 %, air pressure 700 to 1060 hPa
Automatic switch-off:
30 seconds
IP code:
IP 21: Protected against solid foreign particles
with a diameter of more than 12.5 mm, protection
against water droplets
Protection against
electric shock:
Internal power supply, applied part type BF (cuff)