English
English
EN-58
EN-59
Technical information
Technical information
E
E
Cuff:
Cuff type H7, wrist circumference 12.5 cm - 21.5
cm
Operating conditions: Ambient temperature 10°C to 40°C, relative
humidity 15% to 85% non-condensing, air pres-
sure 700 to 1060 hPa
Conditions for storage
and transport:
Ambient temperature -20°C to +60°C
Relative humidity 15% to 95%, non-condensing
Inflation and defla-
tion:
Regulated automatically
IP rating:
IP 20: Protected against solid foreign bodies
with a diameter from 12.5 mm, no protection
against water.
Protection against
electric shock:
Internal power supply, applied part type BF
(cuff)
Operating mode:
Continuous operation
Classification:
Internal power supply using batteries.
4. Original spare parts and accessories
The following original spare parts or accessories are available
through specialist retailers:
Subject to technical modifications.
• Cuff type H7 (for wrist circumference 12.5-21.5 cm)
Part no. 2105601, PZN-13975619
6. Explanation of symbols
This product complies with Council Directive 93/42/
EEC of 5 September 2007 concerning medical devices
and bears the mark CE 0123 (TÜV SÜD Product Service
GmbH).
0123
Important: No modifications, e.g. opening the device (except to
replace the batteries), may be made to this device without the
manufacturer’s permission.
Upon request, responsible authorities or authorised mainte-
nance services receive a “Test instruction for metrological inspec-
tion“ from the manufacturer.
Please only submit the device for metrological inspection togeth-
er with the cuff and the instructions for use.
Degree of protection against electric shock TYPE BF
Device serial number
Consult the instructions for use
Keep dry
5. Metrological inspection
Generally, it is recommended that a metrological inspection is
performed every two years. However, professional users in Ger-
many are obligated to do so in accordance with the “Medical
Products Operator Ordinance” (Medizinprodukte-Betreiberver-
ordnung). This can be performed either by UEBE Medical GmbH,
an authority responsible for metrology or an authorised mainte-
nance service. Please refer to your national regulations.
Reference number = item number
Manufacturer
Protected against solid foreign bodies with a diameter
from 12.5 mm, no protection against water.
IP20