Error
Possible cause
Remedy
Measured values/
pulse values are
extremely high
or low.
Too little rest before
taking the measure-
ment. Movement or
talking during the
measurement.
Repeat the measurement
after a break of 3-5 minutes.
Pay attention to the usage
information on page 42.
Incorrect measure-
ment posture
Repeat the measurement
and maintain the correct
posture. (Page 45)
Measurements
taken in direct
succession
produce different
measured values.
Blood pressure is not
a fixed value. It can go
up or down by more
than 20 mmHg within
a few minutes.
In order to obtain compara-
ble results, always carry out
the measurements under
the same conditions.
Measurement
was interrupted,
the cuff deflated
to 0 and the
measurement
restarted.
Device detects a fault
or that the diastolic
value is too low.
Device shows no fault.
Repeat measurement
Movement during
measurement
If necessary, interrupt mea-
surement and repeat after 5
minutes‘ rest.
Dashes in display
after pressing
Memory button
No measured values
have been stored.
Repeat measurement
The measurement
is interrupted and
the symbol
appears.
Batteries flat
Replace batteries.
The
symbol
appears during
measurement but
the measurement
continues.
Low batteries
You can still perform
approx. 30 measurements,
but should keep spare
batteries available.
No display after
device is switched
on
Batteries inserted
incorrectly
Check position of batteries.
Batteries flat
Replace batteries.
Battery contacts dirty. Clean battery contacts with
a dry cloth.
Date and time are
not visible in the
display
Function was uninten-
tionally deactivated
Activate the date and time
function as described on
page 43.
2. Customer service
Device repairs may only be performed by the manufacturer or an
expressly authorised body. Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Tel.: +49 (0) 9342/924040
Fax.: +49 (0) 9342/924080
Email: [email protected]
Website: www.uebe.com
3. Technical data
Model:
visomat
®
handy express
Size:
88 mm x 64 mm x 26.4 mm
Weight:
approx. 101 g without batteries and cuff
Display:
LCD display (liquid crystal display) for measured
values and check displays
Memory:
2 memories each capable of storing 120
readings
Measurement me-
thod:
Oscillometric determination of systole, diastole
and pulse
Reference method of
the clinical test:
Auscultatory measurement
Pressure display
range:
0-300 mmHg
Measurement range:
Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse measurement: 30-199 pulses/minute
Measurement pre-
cision:
Blood pressure measurement (cuff pressure):
± 3 mmHg
Pulse rate: ± 5%
Power supply:
2 x 1.5 V AAA LR03 alkaline manganese batte-
ries
Power consumption:
2 W