OPM1

REFERENCING SETUP

The LSSM output power is calibrated to -5dBm. If the OPM1 screen displays between -5.00dBm and -7.00dBm, 

press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the dB/dBm button. The OPM1 

screen will display 0.00dB, with some minor variation. DO NOT switch the OPM1 back to the dBm mode.

SETTING REFERENCE

Page 4

M-UM038-01 RP 460 Reference and Test

DMT

Referencing Setup
The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection 

scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 460 and DLS 355. 

The RP 460 should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm 

button once to put it into the dBm mode.

Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm 

and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the 

dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the 

RP 460 back to the dBm mode.

Between -5.00 and -7.00dBm

Hold for 5 seconds

Reference is set

Loss Test
The Test Jumper MUST remain attached to the DLS 355. Do not remove the Test Jumper from the 

optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be cleaned to 

appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper 

into one end of the Fiber-Under-Test using a bulkhead adapter as shown. Plug the other end of the 

Fiber-Under-Test into the LC adapter on the RP 460. The Insertion Loss of the Fiber-Under-Test will be 

shown on the RP 460 screen.

Page 4

M-UM038-01 RP 460 R

efer

ence and T

est

DMT

Refer

encing Setup

The connectors on the T

est Jumper should be cleaned to appropriate standards. Use an i

nspection 

scope to verify fiber cleanliness. Plug the connectors on the T

est Jumper into the RP 460 and DLS 355. 

The RP 460 should be in the dBm measurement mode. If it is not 

in the dBm mode, press the dB/dBm 

button once to put it into the dBm mode.

Setting R

efer

ence

The DLS 355 output power is calibrated to -5dBm. If the RP 460 

screen displays between -5.00dBm 

and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. A

fter 5 seconds, release the 

dB/dBm button. The RP 460 screen will display 0.00dB, with some

 minor variation. DO NO

T switch the 

RP 460 back to the dBm mode.

Betw

een -5.00 and -7.00dBm

Hold for 5 seconds

Refer

ence is set

Loss T

est

The T

est Jumper MUST remain attached to the DLS 355. Do not remove t

he T

est Jumper from the 

optical link. The connectors on the T

est Jumper and the Fiber-Under-T

est should be cleaned to 

appropriate standards. Use an inspection scope to verify fiber c

leanliness. Plug the T

est Jumper 

into one end of the Fiber-Under-T

est using a bulkhead adapter as shown. Plug the other end of th

Fiber-Under-T

est into the L

C adapter on the RP 460. The Insertion Loss of the Fiber-Under-

Test will be 

shown on the RP 460 screen.

OPM1

Between -5.00 and -7.00dBm

Page 4

M-UM038-01 RP 460 Reference and Test

DMT

Referencing Setup
The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection 

scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 460 and DLS 355. 

The RP 460 should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm 

button once to put it into the dBm mode.

Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm 

and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the 

dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the 

RP 460 back to the dBm mode.

Between -5.00 and -7.00dBm

Hold for 5 seconds

Reference is set

Loss Test
The Test Jumper MUST remain attached to the DLS 355. Do not remove the Test Jumper from the 

optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be cleaned to 

appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper 

into one end of the Fiber-Under-Test using a bulkhead adapter as shown. Plug the other end of the 

Fiber-Under-Test into the LC adapter on the RP 460. The Insertion Loss of the Fiber-Under-Test will be 

shown on the RP 460 screen.

Page 4

M-UM038-01 RP 460 Reference and Test

DMT

Referencing Setup
The connectors on the Test Jumper should be cleaned to appropriate standards. Use an inspection 

scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 460 and DLS 355. 

The RP 460 should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm 

button once to put it into the dBm mode.

Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm 

and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the 

dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the 

RP 460 back to the dBm mode.

Between -5.00 and -7.00dBm

Hold for 5 seconds

Reference is set

Loss Test
The Test Jumper MUST remain attached to the DLS 355. Do not remove the Test Jumper from the 

optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be cleaned to 

appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the Test Jumper 

into one end of the Fiber-Under-Test using a bulkhead adapter as shown. Plug the other end of the 

Fiber-Under-Test into the LC adapter on the RP 460. The Insertion Loss of the Fiber-Under-Test will be 

shown on the RP 460 screen.

Hold for 5 seconds

Reference is set

The paired Test Jumper MUST remain attached to the LSSM and OPM1. Do not remove the Test Jumper from 

the optical link. The connectors on the Test Jumper and the fiber-under-test should be cleaned to appropriate 

standards. Use an inspection scope to verify fiber cleanliness. Plug the clipped LC end of the Test Jumper into 

one end of the fiber-under-test using the bulkhead as shown. Plug the loopback on to the other end of the 

fiber-under-test. The Insertion Loss of the fiber-under-test will be shown on the OPM1 screen.

LOSS TEST

Page 8

M-UM038-01 RP 460 Reference and Test

DMT

Referencing Setup
The connectors on the Test Jumper and loopback should be cleaned to appropriate standards. Use 

an inspection scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 

460 and DLS 355. Plug the loopback adapter on to the paired LC side of the Test Jumper. The RP 460 

should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm button 

once to put it into the dBm mode.

Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm 

and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the 

dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the 

RP 460 back to the dBm mode.

Between -5.00 and -7.00dBm

Hold for 5 seconds

Reference is set

Loss Test
The paired Test Jumper MUST remain attached to the DLS 355 and RP 460. Do not remove the Test 

Jumper from the optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be 

cleaned to appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the clipped 

LC end of the Test Jumper into one end of the Fiber-Under-Test using the bulkhead as shown. Plug 

the loopback on to the other end of the Fiber-Under-Test. The Insertion Loss of the Fiber-Under-Test 

will be shown on the RP 460 screen.

Test Jumper

Fiber-Under-Test

Loopback

Page 8

M-UM038-01 RP 460 Reference and Test

DMT

Referencing Setup
The connectors on the Test Jumper and loopback should be cleaned to appropriate standards. Use 

an inspection scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the RP 

460 and DLS 355. Plug the loopback adapter on to the paired LC side of the Test Jumper. The RP 460 

should be in the dBm measurement mode. If it is not in the dBm mode, press the dB/dBm button 

once to put it into the dBm mode.

Setting Reference
The DLS 355 output power is calibrated to -5dBm. If the RP 460 screen displays between -5.00dBm 

and -7.00dBm, press and HOLD the dB/dBm button for 5 seconds. After 5 seconds, release the 

dB/dBm button. The RP 460 screen will display 0.00dB, with some minor variation. DO NOT switch the 

RP 460 back to the dBm mode.

Between -5.00 and -7.00dBm

Hold for 5 seconds

Reference is set

Loss Test
The paired Test Jumper MUST remain attached to the DLS 355 and RP 460. Do not remove the Test 

Jumper from the optical link. The connectors on the Test Jumper and the Fiber-Under-Test should be 

cleaned to appropriate standards. Use an inspection scope to verify fiber cleanliness. Plug the clipped 

LC end of the Test Jumper into one end of the Fiber-Under-Test using the bulkhead as shown. Plug 

the loopback on to the other end of the Fiber-Under-Test. The Insertion Loss of the Fiber-Under-Test 

will be shown on the RP 460 screen.

Paired Test Jumper

Loopback Adapter

The connectors on the Test Jumper and loopback should be cleaned to appropriate standards. Use an

 

inspection scope to verify fiber cleanliness. Plug the connectors on the Test Jumper into the OPM1 and LSSM. 

Plug the loopback adapter on to the paired LC side of the Test Jumper. The OPM1 should be in the dBm 

 

measurement mode. If it is not in the dBm mode, press the dB/dBm button once to put it into the dBm mode.

3330 Earhart Drive, Suite 208  

|

  Carrollton, TX 75006  

|

  Office: 972.556.0916  

|

  Fax: 972.910.8384

PAGE 7

Summary of Contents for TEK-6

Page 1: ...T GUIDE SECTION TWO LSSM LASER LIGHT SOURCE TECHNICAL GUIDE SECTION THREE OPM1 OPTICAL POWER METER QUICK START GUIDE SECTION FOUR OPM1 OPTICAL POWER METER REFERENCE TEST GUIDE SECTION FIVE HFM1 HANDHE...

Page 2: ...SECTION ONE LSSM LASER LIGHT SOURCE QUICK START GUIDE NORTH AMERICA...

Page 3: ...SWIFT This unit offers 1310nm and 1550nm wavelengths When the LSSM is turned on the red light indicates which wavelength is selected The red light blinks to indicate when the 2kHz modulation is turne...

Page 4: ...ts either the 1310nm or 1550nm wavelength on singlemode fiber It is important to use a test jumper to mate the LSSM to the fiber under test TRANSMITTING LIGHT Test Jumper Signal Propagates Down Fiber...

Page 5: ...inserted they can cause damage to the LSSM output port and the unit may need to be repaired OUTPUT PORT MAINTENANCE LSSM When the LSSM s 2kHz modulation is active the selected wavelength indicator wi...

Page 6: ...The LSSM provides over 1000 operations under normal conditions When the LOW BAT LED indicator is illuminated on the device the CR2 battery should be replaced or recharged depending on the battery type...

Page 7: ...long term 15 min warm up 5nm 5nm SC FC ST LC interchangeable 2kHz Push button toggle auto off CR2 10 to 50 C 30 to 60 C 6 1 x 0 94 x 0 75 15 5 x 2 83 x 1 9cm 3oz 85 4g 3330 Earhart Drive Suite 208 Car...

Page 8: ...FCC Rules These limits are designed to provide reasonable protection against harmful inter ference in a residential installation This equipment generates uses and can radiate radio frequency energy a...

Page 9: ...SECTION TWO LSSM LASER LIGHT SOURCE TECHNICAL GUIDE NORTH AMERICA...

Page 10: ...our FI2K Easy removal of the included SC adapter allows the light source ferrule to be cleaned preventing contamination of test jumpers Additional adapters are available This device maintains full co...

Page 11: ...21 CFR 1040 11 Push button toggle auto off CR2 10 to 50 C 30 to 60 C 6 1 x 0 94 x 0 75 15 5 x 2 83 x 1 9cm 3oz 85 4g Interchangable Adapter SC shown Low Battery Indicator Modulation Control 2kHz CW Wa...

Page 12: ...SECTION THREE OPM1 OPTICAL POWER METER QUICK START GUIDE NORTH AMERICA...

Page 13: ...ed to internal storage The display flashes the indexed location where the reading has been saved Toggles through the saved readings on the device The indexed location wavelength and reading will flash...

Page 14: ...ill briefly appear on screen every 8 seconds Press the button to toggle between dB and dBm modes Hold the button down when in dBm mode to set a reference on the OPM1 Press the button once to turn the...

Page 15: ...S Press and hold the Save and Wavelength buttons simultaneously to delete all saved readings in the OPM1 s internal memory The screen flashes 0000 to indicate that the memory is empty NON VOLATILE MEM...

Page 16: ...ckaging is damaged or otherwise unsatisfactory LOW BATTERY The OPM1 provides over 1000 operations under normal conditions When the BATT indicator is shown on the device screen the CR2 battery should b...

Page 17: ...ailable LCD 2kHz 4000 measurements Mini USB live readings or data dump Push button toggle auto off 10 to 55 C 30 to 70 C CR2 6 1 x 0 94 x 0 75 15 5 x 2 83 x 1 9cm 3oz 85 4g 3330 Earhart Drive Suite 20...

Page 18: ...FCC Rules These limits are designed to provide reasonable protection against harmful inter ference in a residential installation This equipment generates uses and can radiate radio frequency energy a...

Page 19: ...SECTION FOUR OPM1 OPTICAL POWER METER REFERENCE TEST GUIDE NORTH AMERICA...

Page 20: ...ch must be taken with OPM1 optical power meter to perform insertion loss testing on a variety of fiber types and setups INTRODUCTION EXAMPLE 1 Singlemode Fiber Point to Point 1 Jumper Page 2 EXAMPLE 2...

Page 21: ...ber under te Pow The 155 wav mat to t ada ada Lig The 155 wav ada 400m Fiber 1m Jumper The fiber under test is singlemode with LC connectors on each end The fiber under test may be up to 250 kilometer...

Page 22: ...8 01 RP 460 Reference and Test DMT Referencing Setup The connectors on the Test Jumper should be cleaned to appropriate standards Use an inspection scope to verify fiber cleanliness Plug the connector...

Page 23: ...da ada Lig The 155 wav ada 400m Fiber 1m Jumper The fiber under test is singlemode with LC connectors on each end This method allows testing of connectors inside bulkheads One or both ends of this fib...

Page 24: ...rence and Test DMT Referencing Setup The connectors on the Test Jumper should be cleaned to appropriate standards Use an inspection scope to verify fiber cleanliness Plug the connectors on the Test Ju...

Page 25: ...400m Fiber 1m Jumper The fiber under test is paired singlemode with paired LC connectors on each end The fiber under test may be up to 250 kilometers long In this example the fiber under test is 400 m...

Page 26: ...e RP 460 screen will display 0 00dB with some minor variation DO NOT switch the RP 460 back to the dBm mode Between 5 00 and 7 00dBm Hold for 5 seconds Reference is set Loss Test The Test Jumper MUST...

Page 27: ...imode with LC connectors on each end In this example the fiber under test is 400 meters long TEST JUMPER The test jumper is a one meter multimode fiber with SC and LC connectors The SC connector will...

Page 28: ...er test will be shown on the OPM1 screen LOSS TEST The connectors on the Test Jumper should be cleaned to appropriate standards Use an inspection scope to verify fiber cleanliness Plug the connectors...

Page 29: ...o 250 kilometers long In this example the fiber under test is 400 meters long TEST JUMPER The test jumper is paired for loopback testing One end has an SC connector and an LC connector and the other e...

Page 30: ...der Test should be cleaned to appropriate standards Use an inspection scope to verify fiber cleanliness Plug the clipped LC end of the Test Jumper into one end of the Fiber Under Test using the bulkhe...

Page 31: ...SECTION FIVE HFM1 HANDHELD INSPECTION MICROSCOPE QUICK START GUIDE NORTH AMERICA...

Page 32: ...hes and contamination with its clear optical performance and integrated laser safety filters and is ideal for examining ferrule polish quality INTRODUCTION Three AAA batteries included ACCESSORIES 2 5...

Page 33: ...STRUCTURE NOTE The two settings on the handheld fiber microscope provide an excellent view of the ferrule end faces We recommend 400x magnification for single mode and 200x magnification for multi mo...

Page 34: ...SECTION SIX VFF1 VISUAL FAULT FINDER QUICK START GUIDE NORTH AMERICA...

Page 35: ...the fibers during spicing operation The choice of a continuous wave output mode for steady fault illumination or a flashing output mode makes for easier fault location INTRODUCTION Two AA batteries in...

Page 36: ...elligent design assists in locating breakpoints of a fiber 3 1 Always vertically insert and remove the fiber connector Otherwise the ceramic tube may break 2 The connector must be well polished and cl...

Reviews: