Important Control Solution Information
Use only
TaiDoc
control solutions with your meter.
Do not use the control solution beyond the expiry date or 3
months after first opening. Write the opening date on the control
solution vial and discard the remaining solution after 3 months.
It is recommended that the control solution test be done at room
temperature (20°C to 25°C/68°F to 77°F). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an ac-
curate result.
Store the control solution tightly closed at temperatures between
2°C to 30°C (35.6°F and 86°F). Do NOT freeze.
Do not use test strips beyond the expiry date. This may cause
inaccurate results.
Do not bend, cut, or alter a test strip in any way.
Keep the strip vial away from children since the cap and the test
strip may be a choking hazard. If swallowed, promptly see a doctor
for help.
For further information, please refer to the test strip package insert.