Problem

Cause

Solution

No power supply;
Fuse damaged;
Power transformer is damaged;

Bad working conditions; Air flow
Something between the scale pan
and working table.
The power exceeds its permissible
value and is unstable;
Static electricity

Improper calibration
The weight of the recipient has
not been tared
Big temperature difference
The balance is not horizontal

Plug in adapter; Change the
fuse;
Change the power transformer;
If problem persists, send the
balance to the Technical
Service for repair.

Improve the working condition,

Close the windshield;

Remove the pan and clean well

the balance surface;

Connect the balance to power

supply 110-220 V AC

Static Elimination

Make calibration Make
tare
Put into suitable
enviroment Adjust level of
balance

Balance

Balance pan

Adapter

Instruction manual

Weight(external

calibration) glove

1

1

1

1

1

1

6. COMPONENTS LIST

No display

Unstable display

Poor accuracy

5.TROUBLESHOOTING

Summary of Contents for 1004

Page 1: ...Analytical Balance Instruction Manual For questions or concerns e mail service ussolid com visit www ussolid com or call 800 243 5428...

Page 2: ...calibration weight value setting linearity calibration 4 OPERATION unit conversion piece counting percent weighing density determination dynamic weighing baud rate print setting sensitivity setting fi...

Page 3: ...quantity of objects Safety Precautions Verify that the AC Adapter input voltage matches the local AC power supply Use the scale only in dry locations For best results use the scale in the specified o...

Page 4: ...r Capacity Electronic Analytical Balance 0 1mg internal calibration 1004 0 100g 1204 0 120g 2004 0 200g 2204 0 220g 3004 0 300g 3204 0 320g 4204 0 420g 5204 0 520g 0 1mg 2s 30 60 mins g oz ct mg 0 1mg...

Page 5: ...its Min weighting Repeatability Linearity Interface Pan size Product size N W Power Model Model Model Model 10035 10035N 10055 10055N 10085 10085N 20035X 20035N 20055 20055N 20085 20085N 100g 30g 100g...

Page 6: ...ours under the power on state to adapt the balance to the new temperature Attention If the number on display is not stable at the first usage it is caused by the temperature difference of the environm...

Page 7: ...from CAL 0 to CAL 200 and flashes put on the 200g weight for FA2204X model display shows 200 0000 remove the weight and finish the calibration internal calibration weight value setting internal calib...

Page 8: ...n procedure Unit conversion under the weighing interface press and hold PRINT to change the unit the default unit g oz ct mg other unit is optional piece counting setting press and hold ON button unde...

Page 9: ...the sample weight the display shows 100 00 If the same contrast is used the second time no re sampling is required Long press PRINT button to exit the percent weighing and return to the weighing mode...

Page 10: ...splay will show the result of density W1 W2 no display press and hold RPINT to exit density result Follow 4 4 setting set the density weighing to dEn 0 press TARE button to exit density determination...

Page 11: ...nter into baud rate setting display shows F press and hold PRINT to set up baud rate the display shows baud rate 1200 baud rate 2400 baud rate 4800 baud rate 9600 press TARE to save the setting and ex...

Page 12: ...ust the sensitivity without consulting the factory Filter level setting press and hold ON button under weighing interface the balance enters into the main menu mode release ON button until the display...

Page 13: ...face CLOC CLOC CLOC 4 OPERATION RS232 interface Connection Balance 9 pins PC Printer 9 pins RXD Input 2 2 TXD Output 3 3 GND Ground 5 5 Output data format Type or da ta data data data data data data d...

Page 14: ...ntal Plug in adapter Change the fuse Change the power transformer If problem persists send the balance to the Technical Service for repair Improve the working condition Close the windshield Remove the...

Reviews: