PAGE 11 OF 39
0318 IH-7095
SPECIFICATIONS
ELECTRICAL SPECIFICATIONS
Accuracy is given as ± (% of r least significant digit) at 73.4ºF ± 9ºF with relative humidity lower than 80%.
Accuracy is specified for a period of one year after calibration.
1. Temperature Coefficient is 0.1 x specified accuracy / ºC, < 18ºC (64.5ºF), > 28ºC (82.4ºF)
2. AC Functionality: ACV and ACA specifications are AC coupled, True RMS, 45 to 400Hz;
For non-sinusoidal waveforms, additional accuracy crest factor (C.F.) considerations exist as detailed below:
NOTE: Duty add 3.0% for C.F. 1.0~2.0; add 5.0% for C.F. 2.0~2.5; add 7.0% for C.F. 2.5~3.0.
FUNCTION
RANGE
RESOLUTION
ACCURACY
DC Voltage
40.00mV
10μV
±(0.7% + 5 digits)
400.0mV
0.1mV
±(0.5% + 2 digits)
4.000V
1mV
±(0.5% + 3 digits)
40.00V
10mV
400.0V
0.1V
600V
1V
Input Impedance: 10M (1G
Ω
for mV range); Overload Protection: AC/DC 600V
AC Voltage
40.00mV
10μV
±(1.0% + 3 digits)
400.0mV
0.1mV
4.000V
1mV
±(0.8% + 3 digits)
40.00V
10mV
400.0V
0.1V
600V
1V
±(1.0% + 3 digits)
LPC
400.0V
0.1V
±(4.0% + 3 digits)
LoZ
600V
------------
±(1.2% + 5 digits)
Maximum input voltage: ±600Vrms.
Input impedance: 10M
Ω
for all ranges (ACV frequency response is 45~400Hz).
Allow the meter to stabilize for 10 minutes after using LoZ (low impedance) mode.
Accuracy stated for 5 to 100% of range.
FUNCTION
DC Current
RANGE
RESOLUTION
ACCURACY
μA
400.0μA
0.1μA
±(0.7% + 2 digits)
4000μA
1μA
mA
40.00mA
10μA
400.0mA
0.1mA
A
4.000A
1mA
±(1.0% + 3 digits)
10.00A
10mA
Overload protection:
μA mA range: (F1) fuse (6.3×32)mm: FF 0.6A H 600V (CE) fast acting
10A range: (F2) fuse (6.3×32)mm: F 10A H 600V (CE) fast acting