© 2021 TytoCare All rights reserved.
74 of 76
TytoPro General Operation User Guide
Appendix D. Electromagnetic Compatibility (EMC) Information
Recommended separation distances between portable and mobile RF communications
equipment and the Tyto Device
Rated
maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
outside ISM
bands
150 kHz to 80 MHz
in ISM bands
80 MHz to 800
MHz
800 MHz to 2,5
GHz
0, 01
0.12
0.2
0.4
1
0, 1
0.37
0.64
1.3
2.6
1
1.17
2
4
8
10
3.7
6.4
13
26
100
11.7
20
40
80
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Warning
: Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
Other equipment could interfere with the Tyto Device, even if the other equipment complies with
CISPR emission requirements.
If the Tyto Device does not respond due to EM phenomena, you can force a shutdown. Press and
hold the physical button on the top of the Device for ten seconds to force the system to shut down
and restart.
Warning
: Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Warning
: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the
TytoCare Device, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.