- 31 -
3.3.9 TRIGger
: TRIGger
Function:
Query or set trigger mode. Or immediately trigger for one time.
Parameter: INTernal (Internal trigger)
EXTernal (External trigger)
IMMediate (Immediately trigger for one time)
Return:
INTERNAL | EXTERNAL
: TRIGger:DELay
Function:
Query or set trigger delay time.
Parameter: 0~6000 (ms)
Return:
0~6000
3.3.10 CORRection
: CORRection
Function:
Clear data.
Parameter: OPEN (Single point open circuit clearing)
OPEN_ALL (All points open circuit clearing)
SHORt (Single point short circuit clearing)
SHORt_ALL (All points short circuit clearing)
3.3.11 COMParator
: COMParator[:STATe]
Function:
Query or set comparator status.
Parameter: ON | OFF | 1 | 0
Return:
1 | 0
: COMParator:Auxiliary BIN
Function:
Query or set auxiliary status.
Parameter: ON | OFF | 1 | 0
Return:
1 | 0
: COMParator:BIN CLEar
Function:
Clear all upper and lower limits.
Parameter: ON | OFF | 1 | 0
Return:
1 | 0
: COMParator:COUNt[:STATe]
Function:
Query or set counter status.
Parameter: ON | OFF | 1 | 0
Return:
1 | 0
: COMParator:COUNt:DATA?
Function:
Query counter number of each range.
Return:
<count numbers of fail range>,<counter number of range 1>,<counter number of range
2>,<counter number of range 3>,<counter number of auxiliary range>