- 34 -
3.3.16 PRINt
Function:
Query or set automatic data transmission function.
Parameter: OFF | 0
ON | 1
Return:
0 | 1
Note:
When automatic data transmission is activated, the instrument automatically transmits
measurement data in the following format.
<A>,<B>,<COMP>
<A>: Measurement data of main parameters.
<B>: Measurement data of secondary parameters.
<COMP>: Measurement data of comparator results 0~5.
3.3.17 COMP
: COMParator: COMP:STATe
Function:
Query or set comparator on/off.
Parameter: ON | OFF | 1 | 0
Return:
1 | 0
: COMParator: COMP: MODE
Function:
Query or set comparator mode.
Parameter: 1 | 0
Return:
1 | 0
Note:
1:
Δ
ABS mode.
: COMParator: COMP: LIMit:NOMinal
Function:
Query or set nominal value of main parameters.
Parameter: < numeric_value > (NR3 format)
Return:
< numeric_value > (NR3 format)
: COMParator: COMP: LIMit:A <HIGH>,<LOW>
Function:
Query or set high or low value of main parameters.
Parameter: <HIGH>,<LOW>
Return:
< numeric_value > (HIGH), < numeric_value > (LOW)
Note:
If there is no high or low limits, the returned value will be “9.9999e+37”.
In comparator mode
△
%, the returned value range is -100~100.
: COMParator: COMP: LIMit:B <HIGH>,<LOW>
Function:
Query or set high or low value of secondary parameters.
Parameter: <HIGH>,<LOW> (NR3 format)
Return:
< numeric_value > (HIGH), < numeric_value > (LOW)
Note:
If there is no high or low limits, the
returned value will be “9.9999e+37”.