MODBUS function code 0x10: write multiple registers
This function code is used to write continuous registers to remote devices (1... 123 registers) block
that specifies the value of the registers written in the request data frame. Data is packaged in two
bytes per register. Response frame return function code, start address and number of registers
written.
Request
N = Register number
Figure11:Write multiple register request frames
Response
N = Register number
Figure 12: write multiple register response frames
The request frame and response frame are illustrated below in two registers that write the values
0x000A and 0x0102 to the start address of 2.
15
Function code
Start Address
Number of input registers
number of bytes
Register values
1 byte
2 byte
2 byte
1 byte
N×2 byte
0x10
2 byte
2 byte
1 byte
N×2 byte
Request Frame
Number Systems
Function code
Start address (high byte)
Start address (low byte)
Input register number (high bytes)
Input register number (low bytes)
number of bytes
Register value (high byte)
Register value (low byte)
Register value (high byte)
Register value (low byte)
(Hexadecimal)
0x10
0x00
0x01
0x00
0x02
0x04
0x00
0x0A
0x01
0x02
Response Frame
Number Systems
Function code
Start address (high byte)
Start address (low byte)
Input register number (high bytes)
Input register number (low bytes)
(Hexadecimal)
0x10
0x00
0x01
0x00
0x02
Function code
Start Address
Register number
1 byte
2 byte
2 byte
0x10
0x0000….0xffff
1…123(0x7B)
Figure 13: Examples of writing multiple register request and response frames
8
Menu structure
If the monitor prompts you to enter the calibration security password, press the
【▼】
key or
【
Ӝ
】
key to
set the calibration security password, and then press the
【
ENT
】
key to confirm the calibration security
password.No initial password here,please enter directly by press
【
ENT
】
key.
The following is the menu structure of this instrument,press
ս
MENU
վ
key to enter menu
setting mode:
Type
Unit
Temp. Sensor
Temperature Offset
Temperature Input
Temperature Unit
Calibration Point 1
Calibration Point 2
Calibration Point 3
Calibration Point 4
Calibration Point 5
Correction
Field Calibration
Offset Adjustment
Slope Adjustment
Status
High/Low Setpoint
Limit Value
Hysteresis
Status
High/Low Setpoint
Limit Value
Hysteresis
Sensor
Temperature
Standard
Solution
Calibration
Field
Calibration
Relay 1
Relay 2
Configure
Calibration
Alarm
Dissloved Ozone
mg/L
ppm
NTC2.252 k
Ω
NTC10 k
Ω
Pt100
Pt1000
0.0000
Automatic
Manual
°C
°F
0.01 (Default, can be modified)
2 (Default, can be modified)
5 (Default, can be modified)
10 (Default, can be modified)
20 (Default, can be modified)
Voltage 1
Voltage 2
Voltage 3
Voltage 4
Voltage 5
ON
OFF
High Alarm
Low Alarm
ON
OFF
High Alarm
Low Alarm