We have made every effort to make this operating manual as short and as clear as
possible.
Nevertheless, if you should have any questions about operation, please contact our
competent service technicians, who are always ready to help. They will be pleased to
assist you.
With the magnetic field meter
List-Magnetik MP-2000
, we offer you a first class,
high-end product, with externally connectable axial and tangential field probes. With
the easy to use device, you can precisely measure magnetic DC and AC fields as well
as pulse fields of all kinds. The range of application ranges from the earth's magnetic
field to a field strength of 40,000 A/cm, switchable in Gauss/Oersted and Tesla. The
built-in data logger, the combined digital and analog display as well as the optional
data transfer via USB, allows flexible use.
Especially at high magnetic field strengths, an absolutely interference-free and pre-
cise measurement is required. For this reason, a microcontroller digitizes and linear-
izes the analog measuring signals of the Hall sensor into the probe in the measuring
probes of the
MP-2000
.
The fast peak value memory allows measurements in pulse fields from 0.1 milli-
seconds. In addition, an oscilloscope for the representation of dynamic magnetic
fields can be connected via the special probe P-T4A with analog output.
A further advantage is that the probe cable can be plugged in on both sides connects
the display unit and the probe and can be easily replaced in the event of a cable
break.
The universally applicable MP-2000 tests for residual magnetism, measures magnetic
fields of all types and locates stray detectors for crack detection.
The device has six display ranges:
DC field: 0-200 A/cm, 200-10,000 A/cm, 1,000-4,000 kA/m
AC field: 0-200 A/cm, 200-10,000 A/cm, 1,000-4,000 kA/m
(1 A/cm = 0.1 kA/m = 1.256 Gauss (Oersted) = 0.1256 mT)
Display range switching takes place automatically.
Applicable measuring probes:
Measuring range up to 20,000 A/cm
Axial probes
P-A2
and
P-W2
(angular axial probe)
Tangential probes
P-T2
and
P-Z2
(flexible reed probe)
Additionally, in measuring range up to 40,000 A/cm
Axial probe
P-A4
Tangential probes
P-T4, P-Z4
and
P-T4A Analog-Output
A calibration standard is available on request for checking the calibration of the
device.