Replacing the Air Filter (ELVC only)
During the cooling phase, of the Liquid w/cooling cycle, compressed air is
uses to maintain pressure inside the chamber. The air filter ensures that the
compressed air is pure and free from contaminants before entering the
chamber. The air filtration is performed by a high quality bacteriological
filter.
The filter is mounted at the right side of the autoclave. .
To replace the filter proceed as follows:
1.
Open the right service door of the autoclave.
2.
Unscrew the fittings (upper and lower fitting) connecting the filter to
the piping.
3.
Replace the filter with a new one and reconnect the fittings.
4.
Close the service door.
Air filter