5.2
Parameters
This directory enables you to see and modify sterilization time parameter
(Dry time is not available on this unit).
1.
Enter the MENU (see sec 5.1)
2.
Using the UP/DOWN keys select the parameters option
3.
Press the START/STOP key to enter that option
4.
The CYCLE PARAMETERS screen is displayed
You can see the parameters of the current cycle
1.
Ster Time: sterilization time
2.
Dry Time: the time of drying stage (N/A for this model)
You can increase or decrease the value of each parameter to reach the
desired value.
NOTE: You can only change the values in the range from the default value
up to 99.
1.
To increase or decrease the sterilization time use the UP and
DOWN keys.
2.
To move the curse from one digit to another press the
START/STOP key
3.
After setting the sterilization time, use the START/STOP key, to go
to “SAVE”. Save the new time by pressing the UP key.
4.
Use the START/STOP key to advance to “EXIT” and press the UP
key to leave this section