38
8.1 To obtain better drying results from the E
and EK models, during the dry stage open
the door partially (as described below) at
an angle of approx. 20-25 degrees. The
heaters will be on at a lower power level
and water vapor and humidity will be
evacuated from the chamber efficiently.
Without moving the closing device from
its position, unscrew the closing device as
far as it will go. Allow the door to open to
the maximum that the closing device will
allow (see picture).
9.
At the end of the cycle a buzzer rings for approximately 5 seconds, the
START light switches OFF. The air outlet valve is opened to prevent
formation of a vacuum.
In the event of a program failure, the exhaust valve is opened to release
pressure from the chamber and a continuous buzz will sound for 5 seconds
followed by an interrupted buzz of 7 seconds.
10. Open the door and unload the sterilized material from the chamber.
11. The sterility of instruments processed in unwrapped cycles cannot be
maintained if exposed to a non-sterile environment.
Note:
A minimum time interval of 10 minutes needs to be observed between the
end of a cycle and the start of a new cycle, to prevent overheating of the
autoclave.
Do not touch the strainer’s cover, mounted on the exhaust line, during and
shortly after operation, it will get very hot.
Touching the hot strainer’s cover may cause severe injuries.