19
(9) DN
Pressing these keys in combination with
TEMP
. (5),
STE
TIME
(6),
DRY
Time (7) and
CLOCK
(9) lowers these
values.
(10) CLOCK
Pressing the CLOCK programming key displays the date, with
the cursor under the day. Pressing the UP or DN keys changes
the date. Pressing the CLOCK key again moves the cursor to
the month, then year and then time (hour, minute, second). At
this point the display shows the currently set date and time. If
no key is pressed during a 10-second interval, the system exits
the clock-programming mode and returns to the current
program display.
(11) Water Inlet
Pressing this key continuously, allows for manual filling of the
chamber with water. Once the key is released the water pump
stops, the fill valve closes and water stops entering the chamber.
(12) START
Pressing this key starts the sterilization (or DRY ONLY)
process according to the selected program. Water flows
automatically into the chamber, HEAT and STE. stages
commence and the respective LED indicator lights up.
On completion of the sterilization program the Exhaust stage
will automatically begin, at the end of which a Drying stage (if
previously programmed) will initiate. The respective LED
indicators light up, indicating which stage is operating. Once all
stages have been completed, the final indicator extinguishes and
the screen displays the “Cycle Finished” message.
The process will not start if:
The door is not closed and the
DOOR CLOSED
indicator is off.
The “DOOR UNLOCK” message is displayed.
Note:
Due to inherent elasticity of the door gasket
,
the
CLOSE DOOR indicator may be illuminated green
before a complete seal is made between the door
and the chamber. Therefore, in order to ensure the
door is fully sealed, tighten the door bolt unti
l
‘hand
tight’. Do not overtighten the bol
t
as this may result
in damage to the gasket. Should the autoclave fail
to reach sterilizing temperature/pressure, always
check first if the door is fully sealed. If not, tighten
the
door further, as described above, until
completely sealed.
There is not enough water in the reservoir, (the red
ADD WATER indicator lights and the “ADD
WATER” message is displayed).