Aqua
Fluor
®
U
ser’s Manual
22
Application Specifications
P/N
Application
Sol. Std.
10x10 mm
Cuvettes
12 mm
Vials
Temperature
Coefficients
8000-402
Ammonium
*see note below
Not Available
7000-959
10-029A
Not Available
8000-401
CDOM/FDOM
Not Available
7000-959
10-029A
Not Available
8000-407
Chl
a
Extracted - Acidification
8000-952
Not Available
10-029A
0.3/°C Linear
8000-406
Chl
in vivo
8000-952
7000-959
7000-957
10-029A
1.4%/
o
C
Linear
8000-405
Fluorescein Dye
8000-951
7000-959
7000-957
10-029A
0.0036/°C
Exponential
8000-403
Optical Brighteners for
Wastewater Monitoring
Not Available
7000-959
10-029A
Not Available
8000-412
Phycocyanin
**see note below
(Freshwater Cyanobacteria)
8000-952
7000-959
7000-957
10-029A
Not Available
8000-411
Phycoerythrin
(Marine Cyanobacteria)
8000-952
7000-959
7000-957
10-029A
Not Available
8000-409
Rhodamine Dye
8000-952
7000-959
7000-957
10-029A
0.026/°C
Exponential
8000-408
Turbidity
Not Available
7000-957
10-029A
Not Available
7000-959 = Methacrylate 10x10 mm cuvettes
7000-957 = Polystyrene 10x10 mm cuvettes
10-029A = Glass 12x75 mm test tubes
*Readings on the Ammonium channel using plastic cuvettes have a maximum range of
approximately 10 µM. Minicells are required to achieve readings of 10 µM - 100 µM. Minicell
Adapter P/N 8000-936 and Minicells (500 ct.) P/N 138-0140 are included with Ammonium Channel.
**Readings on the Phycocyanin channel using cuvettes or vials have a maximum range of
approximately 3,500 ppb. Minicells P/N 7000-950 and Minicell adaptor P/N 8000-936 are required to
achieve maximum range of approximately 20,000 ppb.